Design and Performance of a New Thin‐Layer Radial‐Flow Holder for a Quartz Crystal Resonator of an Electrochemical Quartz Crystal Microbalance
- 15 November 2006
- journal article
- Published by Wiley in Electroanalysis
- Vol. 18 (22) , 2168-2173
Abstract
A new compact holder for either 5‐ or 10‐MHz AT‐cut quartz crystal resonator of an electrochemical quartz crystal microbalance was designed, fabricated and characterized. The holder is a hydrodynamically controlled thin‐layer radial‐flow microelectrochemical cell. Its unique feature consists of (i) a micrometer‐screw adjustable distance between the movable coaxial assembly of the Ag/Ag+ pseudoreference electrode and the inlet capillary nozzle with respect to the metal‐film working electrode of the quartz crystal resonator, and (ii) a U‐clamp mountable resonator, easily accessible for change without using any tools. The inlet solution stream is centered axially against the working electrode. The holder performance was tested under different flow conditions. These include hydrodynamic voltammetry measurements on the Fe(CN)$\rm{ {_{6}^{4-}}}$/Fe(CN)$\rm{ {_{6}^{3-}}}$ couple, i.e., a redox system with no mass transfer across the solution–electrode interface, as well as simultaneous chronoamperometry and chronoelectrogravimetry measurements under flow injection analysis (FIA) conditions on the Ag/Ag+ couple, i.e., a system with electrodeposition of a rigid metallic film. Moreover, simultaneous changes of resonant frequency and dynamic resistance were measured under FIA conditions for a glycerol solution, i.e., an electroinactive viscous medium. For the 30<Fm−1 volume flow rate of solution and 50<dFm=30 μL min−1 and d=100 μm, both mass and charge conversion accompanying silver electrodeposition is appreciably high and close to 35%. Simultaneous measurements of the resonant frequency change and current‐potential or current‐time transients allowed investigations of electrochemical processes involving mass changes of rigid deposits while those of the frequency change and dynamic resistance change involve changes of viscoelastic properties of medium.Keywords
This publication has 27 references indexed in Scilit:
- High resolution gravimetric, optical and electrochemical investigations of microbial biofilm formation in aqueous systemsElectrochimica Acta, 2003
- Dependence of the reliability of electrochemical quartz-crystal nanobalance mass responses on the calibration constant, Cf : analysis of three procedures for its determinationElectrochimica Acta, 2003
- Study of the scale inhibition by HEDP in a channel flow cell using a quartz crystal microbalanceElectrochimica Acta, 2001
- Characterization of modified electrodes by electrochemical quartz crystal microbalance, radiotracer technique and impedance spectroscopyElectroanalysis, 1995
- Quartz Crystal Electrogravimetry with Controlled Hydrodynamics: Applications to the Study of Nickel ElectrodepositionJournal of the Electrochemical Society, 1994
- An improved holder for the electrochemical quartz crystal microbalance and its cyclic voltammetry characteristicsElectroanalysis, 1993
- Electroanalytical voltammetry in flowing solutionsAnalytica Chimica Acta, 1986
- Theoretical and practical limitations on the optimization of amperometric detectorsAnalytical Chemistry, 1984
- Electrochemical detectors in liquid chromatography. A short review of detector designIndustrial & Engineering Chemistry Product Research and Development, 1981
- Verwendung von Schwingquarzen zur W gung d nner Schichten und zur Mikrow gungThe European Physical Journal A, 1959