New type of radial collimator for strain measurements by neutron diffraction
Authors:
- 1 January 1995
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (1) , 32-37
Abstract
We present a new type of radial collimator made of boron fibers. Coupled with a position sensitive detector system it is optimal for neutron diffraction studies of internal strain measurements. We show the results of the calibration and first test measurements on such a system. The obtained characteristics are excellent and allow a substantial (30) gain factor in data acquisition time.Keywords