High‐resolution human cervical spinal cord imaging at 7 T
- 20 December 2011
- journal article
- research article
- Published by Wiley in NMR in Biomedicine
- Vol. 25 (7) , 891-899
Abstract
We present high‐resolution anatomical imaging of the cervical spinal cord in healthy volunteers at the ultrahigh field of 7 T with a prototype four‐channel radiofrequency coil array, in comparison with 3‐T imaging of the same subjects. Signal‐to‐noise ratios at both field strengths were estimated using the rigorous Kellman method. Spinal cord cross‐sectional area measurements were performed, including whole‐cord measurements at both fields and gray matter segmentation at 7 T. The 7‐T array coil showed reduced sagittal coverage, comparable axial coverage and the expected significantly higher signal‐to‐noise ratio compared with equivalent 3‐T protocols. In the cervical spinal cord, the signal‐to‐noise ratio was found by the Kellman method to be higher by a factor of 3.5 with the 7‐T coil than with standard 3‐T coils. Cervical spine imaging in healthy volunteers at 7 T revealed not only detailed white/gray matter differentiation, but also structures not visualized at lower fields, such as denticulate ligaments, nerve roots and rostral–caudal blood vessels. Whole‐cord cross‐sectional area measurements showed good agreement at both field strengths. The measurable gray/white matter cross‐sectional areas at 7 T were found to be comparable with reports from histology. These pilot data demonstrate the use of higher signal‐to‐noise ratios at the ultrahigh field of 7 T for significant improvement in anatomical resolution of the cervical spinal cord, allowing the visualization of structures not seen at lower field strength, particularly for axial imaging. Copyright © 2011 John Wiley & Sons, Ltd.Keywords
This publication has 36 references indexed in Scilit:
- A radiofrequency coil configuration for imaging the human vertebral column at 7TJournal of Magnetic Resonance, 2011
- Image‐guided radio‐frequency gain calibration for high‐field MRINMR in Biomedicine, 2009
- Magnetic resonance imaging in acute cervical spinal cord injury: a correlative study on spinal cord changes and 1 month motor recoverySpinal Cord, 2008
- High b‐value q‐space diffusion‐weighted MRI of the human cervical spinal cord in vivo: Feasibility and application to multiple sclerosisMagnetic Resonance in Medicine, 2008
- Morphology and Morphometry of Human Chronic Spinal Cord Injury Using Diffusion Tensor Imaging and Fuzzy LogicAnnals of Biomedical Engineering, 2007
- Line scan diffusion tensor MRI at low magnetic field strength: Feasibility study of cervical spondylotic myelopathy in an early clinical stageJournal of Magnetic Resonance Imaging, 2005
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- Analysis of the diffusion weighted MR microscopy data of excised spinal cord of a rat on the basis of the model of restricted diffusionSolid State Nuclear Magnetic Resonance, 2004
- Gross quantitative measurements of spinal cord segments in humanSpinal Cord, 2004
- Two‐dimensional selective adiabatic pulsesMagnetic Resonance in Medicine, 1992