A Comparative Study of the Performance of Different Corneal Topographers on Children with Respect to Orthokeratology Practice
- 1 May 2005
- journal article
- research article
- Published by Wiley in Optometry and Vision Science
- Vol. 82 (5) , 420-427
Abstract
Purpose.: The purpose of this study was to evaluate and compare the performance of three topographers, Medmont E300, Keratron Scout, and Humphrey Atlas 991, for the measurement of corneal parameters of interest in orthokeratology practice on young children. Methods.: Apical radius (R 0 ), flattest corneal curvature (flat k), eccentricity (e) value, and elevation at 9‐mm chord were measured twice on 22 healthy subjects of mean ± standard deviation (SD) age 11.2 ± 2.2 years using the topographers under investigation. Results.: The repeatability of the Medmont and Humphrey were good for the measurements of R 0 and flat k. The repeatability of the Humphrey was also good for the measurement of e‐value. However, for both topographers, the measurement of elevation was not precise enough for orthokeratology empirical lens design. The Keratron showed poor repeatability for all the parameters evaluated. The agreement of the parameters of interest between these instruments was poor. Conclusions.: The Medmont and the Humphrey showed good repeatability on children, but the precision of elevation measurements was not good enough to be used for designing empiric orthokeratology lenses on children. The Keratron is not recommended to be used on children in orthokeratology fitting and management. Interinstrumental agreements were significantly different and therefore measurements from different topographers should not be used interchangeably.Keywords
This publication has 9 references indexed in Scilit:
- The Performance of Four Different Corneal Topographers on Normal Human Corneas and Its Impact on Orthokeratology Lens FittingOptometry and Vision Science, 2002
- The Accuracy and Precision Performance of Four Videokeratoscopes in Measuring Test SurfacesOptometry and Vision Science, 2000
- The reproducibility of videokeratoscope measurements as applied to the human corneaContact Lens and Anterior Eye, 1999
- Reconstruction of the Corneal Shape with the MasterVue Corneal Topography SystemOptometry and Vision Science, 1997
- Height Measurement of Astigmatic Test Surfaces by a Keratoscope That Uses Plane Geometry Surface ReconstructionAmerican Journal of Ophthalmology, 1996
- Repeatability of Corneal Topography: The "Corneal Field"Journal of Refractive Surgery, 1995
- Characterization of the Inherent Error in a Spherically-Biased Corneal Topography System in Mapping a Radially Aspheric Surface/Comment: Robert B. Mandell, OD, PhD Berkeley, Calif/Comment: Raymond A. Applegate, OD, PhD San Antonio, Tex/Response: Cynthia Roberts, PhDJournal of Refractive Surgery, 1994
- Computer-Assisted Corneal Topography: Accuracy and Reproducibility of the Topographic Modeling SystemJournal of Refractive Surgery, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986