Assessment of accuracy of laboratory testing results, relative to peer group consensus values in external quality control, by bivariate z-score analysis: the example of D-Dimer
Authors:
Open Access
- 19 February 2024
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 62 (8) , 1548-1556
Abstract
Objectives: The aim of this study is to develop a practical method for bivariate z-score analysis which can be applied to the survey of an external quality assessment programme. Methods: To develop the bivariate z-score analysis, the results of four surveys of the international D-Dimer external quality assessment programme of 2022 of the ECAT Foundation were used. The proposed methodology starts by identifying the bivariate outliers, using a Supervised Sequential Hotelling T2 control chart. The outlying data are removed, and all the remaining data are used to provide robust estimates of the parameters of the assumed underlying bivariate normal distribution. Based on these estimates two nested homocentric ellipses are drawn, corresponding to confidence levels of 95 and 99.7 %. The bivariate z-score plot described provides the laboratory with an indication of both systematic and random deviations from zero z-score values. The bivariate z-score analysis was examined within survey 2022-D4 across the three most frequently used methods. Results: The number of z-score pairs included varied between 830 and 857 and the number of bivariate outliers varied between 20 and 28. The correlation between the z-score pairs varied between 0.431 and 0.647. The correlation between the z-score pairs for the three most frequently used varied between 0.208 and 0.636. Conclusions: The use of the bivariate z-score analysis is of major importance when multiple samples are distributed around in the same survey and dependency of the results is likely. Important lessons can be drawn from the shape of the ellipse with respect to random and systematic deviations, while individual laboratories have been informed about their position in the state-of-the-art distribution and whether they have to deal with systematic and/or random deviations.Keywords
This publication has 27 references indexed in Scilit:
- Improving Youden plots by including analytical performance specificationsClinica Chimica Acta, 2022
- Diagnostic accuracy of four different D-dimer assays: A post-hoc analysis of the YEARS studyThrombosis Research, 2021
- The D‐dimer assayAmerican Journal of Hematology, 2019
- Appraising laboratory quality and value: What's missing?Clinical Biochemistry, 2017
- Harmonisation of D-dimer — A call for actionThrombosis Research, 2015
- Comparison of traditional, trimmed traditional and robust Youden chartsClinica Chimica Acta, 2015
- A study on the utilization of the Youden plot to evaluate proficiency test resultsAccreditation and Quality Assurance, 2013
- Multicenter evaluation of a new quantitative highly sensitive D-dimer assay, the Hemosil® D-dimer HS 500, in patients with clinically suspected venous thromboembolismThrombosis Research, 2010
- The role of proficiency testing in achieving standardization and harmonization between laboratoriesClinical Biochemistry, 2009
- Linking analytic performance goals to medical outcomeClinica Chimica Acta, 1997