Validation of a micro‐CT technique for measuring volumetric wear in retrieved acetabular liners
Authors:
- A. E. BowdenABA. E. Bowden
- A. A. EdidinAEA. A. Edidin
- 21 July 2005
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 75B (1) , 205-209
Abstract
In this study, a novel micro‐CT–based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high‐resolution micro‐CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16‐bit precision) with the use of a 1024 × 1024 in‐plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm3/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra‐ and interobserver uncertainty analysis showed the method to be both accurate and repeatable. © 2005 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2005Keywords
This publication has 25 references indexed in Scilit:
- Analysis of a retrieved UHMWPE acetabular cup crosslinked in air with 1000 kGy of gamma radiationJournal of Orthopaedic Research, 2004
- Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements2004
- Polyethylene cross‐linking by two different methods reduces acetabular liner wear in a hip joint wear simulatorJournal of Orthopaedic Research, 2003
- COMPARISON OF TWO AND THREE-DIMENSIONAL COMPUTERIZED POLYETHYLENE WEAR ANALYSIS AFTER TOTAL HIP ARTHROPLASTYJournal of Bone and Joint Surgery, 2003
- The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: role of the third-body wear processBiomaterials, 2002
- Comparison of polyethylene wear with femoral heads of 22 mm and 32 mm. A prospective, randomised study.The Journal of Bone and Joint Surgery. British volume, 2002
- Wear of conventional and cross‐linked ultra‐high‐molecular‐weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulatorJournal of Biomedical Materials Research, 2002
- A Clinically Practical Method of Manually Assessing Polyethylene Liner ThicknessJournal of Bone and Joint Surgery, 2001
- Radiographic wear measurements in a cementless metal-backed modular cobalt-chromium acetabular componentThe Journal of Arthroplasty, 2001
- Rate of Wear in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975