A depth-encoding PET detector that uses light sharing and single-ended readout with silicon photomultipliers
Authors:
- 1 February 2018
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 63 (4) , 045009
Abstract
Detectors with depth-encoding capability and good timing resolution are required to develop high-performance whole-body or total-body PET scanners. In this work, depth-encoding PET detectors that use light sharing between two discrete crystals and single-ended readout with silicon photomultipliers (SiPMs) were manufactured and evaluated. The detectors consisted of two unpolished 3 × 3 × 20 mm 3 LYSO crystals with different coupling materials between them and were read out by Hamamatsu 3 × 3 mm 2 SiPMs with one-to-one coupling. The ratio of the energy of one SiPM to the total energy of two SiPMs was used to measure the depth of interaction (DOI). Detectors with different coupling materials in-between the crystals were measured in the singles mode in an effort to obtain detectors that can provide good DOI resolution. The DOI resolution and energy resolution of three types of detector were measured and the timing resolution was measured for the detector with the best DOI and energy resolution. The optimum detector, with 5 mm optical glue, a 9 mm triangular ESR and a 6 mm rectangular ESR in-between the unpolished crystals, provides a DOI resolution of 2.65 mm, an energy resolution of 10.0% and a timing resolution of 427 ps for events of E > 400 keV. The detectors simultaneously provide good DOI and timing resolution, and show great promise for the development of high-performance whole-body and total-body PET scanners.Keywords
Funding Information
- Hundred-Talent Program of the Chinese Academy of Sciences (Yongfeng Yang)
- National Natural Science Foundation of China (11575285)
- Shenzhen Peacock Plan (KQTD2016053117113327)
- the Natural Science Foundation of Guangdong Province (2014A030312006)
- the Basic Research Program of Shenzhen (JCYJ20160608153434110)
This publication has 44 references indexed in Scilit:
- Total-body imaging: Transforming the role of positron emission tomographyScience Translational Medicine, 2017
- Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operationPhysics in Medicine & Biology, 2016
- A 32 mm × 32 mm × 22 mm monolithic LYSO:Ce detector with dual-sided digital photon counter readout for ultrahigh-performance TOF-PET and TOF-PET/MRIPhysics in Medicine & Biology, 2016
- A combined time‐of‐flight and depth‐of‐interaction detector for total‐body positron emission tomographyMedical Physics, 2016
- Physical Performance of the new hybrid PET/CT Discovery-690Medical Physics, 2011
- PET systems: the value of added lengthEuropean Journal of Nuclear Medicine and Molecular Imaging, 2010
- Sensitivity assessment of wide Axial Field of View PET systems via Monte Carlo simulations of NEMA-like measurementsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Performance of a high sensitivity PET scanner based on LSO panel detectorsIEEE Transactions on Nuclear Science, 2006
- Neural network-based position estimators for PET detectors using monolithic LSO blocksIEEE Transactions on Nuclear Science, 2004
- The effect of camera geometry on singles flux, scatter fraction and trues and randoms sensitivity for cylindrical 3D PET-a simulation studyIEEE Transactions on Nuclear Science, 2000