Forward-Viewing Intravascular Ultrasound: Design and Fabrication for High Sensitivity 3D Imaging of Hemodynamics
- 1 August 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 73 (8) , 2899-2910
Abstract
Percutaneous coronary intervention (PCI) treats myocardial ischemia by stenting stenotic coronary arteries. Wall shear stress (WSS) is an established indicator of plaque rupture risk and restenosis likelihood, yet is not currently used to guide stenting. To address this gap, we developed a forward-viewing intravascular ultrasound array and demonstrated 3D hemodynamic imaging in coronary artery phantoms. A 3 mm, 172 element forward-viewing 2D array was designed, simulated, and fabricated. Field II was used to simulate aperture performance, and finite element modeling was conducted to investigate element geometry effects on vibrational modes. The array was characterized electrically and acoustically, then used to acquire 3D color flow images in coronary artery-mimicking phantoms using plane wave compounding, SVD-based wall filtering, transverse oscillation velocity estimation, and Fourier sine series WSS estimation. The fabricated array achieved a center frequency of 31 MHz, a -6 dB bandwidth of 60%, and a single-element SNR of 20 ± 1 dB, with 164 of 172 elements operational. Lateral and axial resolutions of 170 μm and 150 μm, respectively, were measured at a depth of 5 mm. Peak velocities of 1 cm/s and 2.6 cm/s were measured in straight and stenotic phantoms, and WSS was derived from these 3D images. 3D finite element modeling provided new insights into element geometry effects on vibrational modes for IVUS 2D arrays, enabling a 31 MHz center frequency, 60% bandwidth, and 95% yield. This work represents the first demonstration of 3D imaging of WSS using a forward-viewing IVUS 2D array, a step toward IVUS-based hemodynamic guidance.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health (R01EB031101)
- National Science Foundation (ECCS-2025462)
- Georgia Tech Institute for Electronics and Nanotechnology
- National Nanotechnology Coordinated Infrastructure
This publication has 47 references indexed in Scilit:
- A Pseudo-Spectral Method for Wall Shear Stress Estimation from Doppler Ultrasound Imaging in Coronary ArteriesCardiovascular Engineering and Technology, 2024
- Transrectal ultrasound vector projectile imaging for time‐resolved visualization of flow dynamics in the male urethra: A clinical pilot studyMedical Physics, 2023
- Dynamic Coronary Blood Flow Velocity and Wall Shear Stress Estimation Using Ultrasound in an Ex Vivo Porcine HeartCardiovascular Engineering and Technology, 2023
- Coronary In-Stent RestenosisJACC, 2022
- Multifrequency ultrasonic transducers based on dual vibration and harmonic modeSensors and Actuators A: Physical, 2022
- Improving PCI Outcomes Using Postprocedural Physiology and Intravascular ImagingJACC: Cardiovascular Interventions, 2021
- Role of Low Endothelial Shear Stress and Plaque Characteristics in the Prediction of Nonculprit Major Adverse Cardiac EventsJACC: Cardiovascular Imaging, 2018
- In vivo blood flow and wall shear stress measurements in the vitelline networkExperiments in Fluids, 2008
- Can we afford to eliminate restenosis?JACC, 2004
- Alterations of phasic coronary artery flow velocity humans during percutaneous coronary angioplastyJACC, 1992