Efficacy and safety of lower versus higher CO2extraction devices to allow ultraprotective ventilation: secondary analysis of the SUPERNOVA study

Authors:

Abstract
Retrospective analysis of the SUPERNOVA trial exploring the hypothesis that efficacy and safety of extracorporeal carbon dioxide removal (ECCO2R) to facilitate reduction of tidal volume (VT) to 4 mL/kg in patients with acute respiratory distress syndrome (ARDS) may differ between systems withlower(area of membrane length 0.59 m2; blood flow 300–500 mL/min) andhigher(membrane area 1.30 m2; blood flow between 800 and 1000 mL/min) CO2extraction capacity. Ninety-five patients with moderate ARDS were included (33 patients treated withlowerand 62 patients treated withhigherCO2extraction devices). We found that (1) VTof 4 mL/kg was reached by 55% and 64% of patients with the lower extraction versus 90% and 92% of patients with higher extraction devices at 8 and 24 hours from baseline, respectively (p2R-related haemolysis and bleeding was higher withlowerthan withhigherextraction devices (21% vs 6%, p=0.045% and 27% vs 6%, p=0.010, respectively). Although VTof 4 mL/kg could have been obtained with all devices, this was achieved frequently and with a lower rate of adverse events by devices withhigherCO2extraction capacity.
Funding Information