A simple, low-cost, computerized, very low-frequency lock-in amplifier is described for signal-to-noise enhancement. This system is based on a low-cost digital multimeter connected to a microcomputer through an IEEE-488 interface. The system works with signals that can be modulated at ≃1 Hz. The signal-to-noise ratio, calculated in real time during data acquisition, can be used to terminate acquisition either after the desired signal-to-noise ratio is reached or after a specified number of samples has been averaged. The system is also suitable for eliminating large constant backgrounds on small signals. Possible applications include spectrophotometry, spectrofluorimetry, and ac calorimetry.Keywords