Characterization and Performance of PADME's Cherenkov-Based Small-Angle Calorimeter

Authors:
  • AFA. Frankenthal
    A. Frankenthal
  • JAJ. Alexander
    J. Alexander
  • BBB. Buonomo
    B. Buonomo
  • ECE. Capitolo
    E. Capitolo
  • CCC. Capoccia
    C. Capoccia
  • CCC. Cesarotti
    C. Cesarotti
  • RSR. De Sangro
    R. De Sangro
  • CGC. Di Giulio
    C. Di Giulio
  • FFF. Ferrarotto
    F. Ferrarotto
  • LFL. Foggetta
    L. Foggetta
  • GGG. Georgiev
    G. Georgiev
  • PGP. Gianotti
    P. Gianotti

Abstract
The PADME experiment, at the Laboratori Nazionali di Frascati (LNF), in Italy, will search for invisible decays of the hypothetical dark photon via the process $e^+e^-\rightarrow \gamma A'$, where the $A'$ escapes detection. The dark photon mass range sensitivity in a first phase will be 1 to 24 MeV. We report here on measurement and simulation studies of the performance of the Small-Angle Calorimeter, a component of PADME's detector dedicated to rejecting 2- and 3-gamma backgrounds. The crucial requirement is a timing resolution of less than 200 ps, which is satisfied by the choice of PbF$_2$ crystals and the newly released Hamamatsu R13478UV photomultiplier tubes (PMTs). We find a timing resolution of 81 ps (with double-peak separation resolution of 1.8 ns) and a single-crystal energy resolution of 5.7%/$\sqrt{E}$ with light yield of 2.07 photo-electrons per MeV, using 100 to 400 MeV electrons at the Beam Test Facility of LNF. We also propose the investigation of a two-PMT solution coupled to a single PbF$_2$ crystal for higher-energy applications, which has potentially attractive features.