Near vision restoration with refractive lens exchange and pseudoaccommodating and multifocal refractive and diffractive intraocular lenses
Authors:
- Jorge L. AlióJAJorge L. Alió
- Marco TavolatoMTMarco Tavolato
- Fernando De la HozFHFernando De la Hoz
- Pascual ClaramontePCPascual Claramonte
- José-Luis Rodríguez-PratsJRJosé-Luis Rodríguez-Prats
- Ahmed GalalAGAhmed Galal
- 1 December 2004
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 30 (12) , 2494-2503
Abstract
To evaluate near visual performance after implantation of a pseudoaccommodating intraocular lens (IOL) (Crystalens AT-45, eyeonics) or a multifocal IOL (refractive model, AMO Array; diffractive model, AcriTec TwinSet) after lens surgery. Instituto Oftalmologico de Alicante, Miguel Hernandez University, Alicante, Spain. Forty patients were included in this prospective clinical comparative study. The patients were divided into 3 groups according to the type of IOL implanted. Group 1 included 24 eyes (12 patients) implanted with the Crystalens IOL; Group 2, 32 eyes (16 patients) with the Array IOL, and Group 3, 24 eyes (12 patients) with the TwinSet IOL. Bilateral phacoemulsification and IOL implantation were performed in all patients, and the follow-up was 1 year. The postoperative main visual outcome evaluations were uncorrected and best corrected distance and near visual acuities, mean add for near, and best distance-corrected near acuity. In Group 1, the mean uncorrected near acuity was 20/40 preoperatively and 20/25 after 1 year. The mean preoperative best distance-corrected near acuity was 20/32 and 20/25 after 1 year. In Group 2, the mean uncorrected near acuity was 20/40 and 20/25, respectively. The mean preoperative best distance-corrected near acuity was 20/32 and 20/25 after 1 year. In Group 3, the mean uncorrected near acuity was 20/63 preoperatively; after 1 year, it was approximately 20/25. The mean best distance-corrected near acuity was 20/50 and 20/25, respectively. A neodymium:YAG laser capsulotomy was performed for posterior capsule opacification when required, with no change in the refractive outcome. Implantation of multifocal and pseudoaccommodating IOLs provides adequate near vision restoration. The TwinSet IOL provided faster recovery of near vision than the other 2 IOLs. The Crystalens IOL provided less postoperative visual phenomena with favorable near vision. The Array IOL achieved the most comfortable distance and near vision.Keywords
This publication has 17 references indexed in Scilit:
- Distance and near contrast sensitivity function after multifocal intraocular lens implantationJournal of Cataract & Refractive Surgery, 2003
- Multifocal intraocular lens implantation in prepresbyopic patients with unilateral cataractOphthalmology, 2002
- Subjective photic phenomena with refractive multifocal and monofocal intraocular lensesJournal of Cataract & Refractive Surgery, 2001
- Cataract extraction with multifocal intraocular lens implantationOphthalmology, 2000
- Scleral expansion surgery does not restore accommodation in human presbyopiaOphthalmology, 1999
- Ringwulstlinse mit Zoomwirkung zur Verstärkung einer Pseudoakkommodation und deren Erklärung aus erweiterter AkkommodationstheorieSpektrum der Augenheilkunde, 1997
- Experience with an Accommodating IOLJournal of Cataract & Refractive Surgery, 1996
- Comparison of a diffractive bifocal and a monofocal intraocular lensJournal of Cataract & Refractive Surgery, 1996
- Small-diameter corneal inlay in presbyopic or pseudophakic patientsJournal of Cataract & Refractive Surgery, 1995
- Diffractive multifocal intraocular lens implants for unilateral cataracts in prepresbyopic patients.British Journal of Ophthalmology, 1992