Construction of an Artificial Heart Pump Performance Test System

Authors:

Abstract
A hydraulic loop, which simulates pressure/flow response of the human circulatory system, is needed to bench test the various versions of rotary left ventricular assist devices (LVADs). This article describes the design of such a loop and the simulated response of different physiological states, such as a healthy person in sleep, rest, and mild physical activity, and in different pathological states. The loop consists of: (1) pulsatile left and right cardiac simulators; (2) air/water tanks to model the venous and arterial compliances; (3) tygon tubes to model the venous, arterial, and other system flow resistances; and (4) a tuning clamp to model the variation in system resistance characteristics under different cardiac pressure/flow conditions. The simulated responses were compared to the data found in the literature to validate the loop performance prior to LVAD testing.

This publication has 12 references indexed in Scilit: