Abstract
For accurately detecting capacitance changes of the differential‐capacitance transducer, which is widely used in environmental and chemical engineering, a low‐cost interface circuit was developed. The circuit uses a “lock‐in detection” circuit with feedback loop containing a DC amplifier and a modulator to yield such advantages over the others that it reduces the sensitivity to element parameter variations in the “lock‐in detection” circuit, that it decreases the sensitivity to the disturbances within closed‐loop, and that it is immune to parasitic capacitance, resistance, amplitude, and frequency of the excitation voltage source. The circuit has the capability to detect a capacitance change as small as 0.0078% of the total capacitance, up to 80 pF. A capacitance diaphragm pressure transducer and a standard pressure source (Rusca 2465 piston gauge) are connected to the interface to confirm the analysis.

This publication has 4 references indexed in Scilit: