Simultaneous measurements of intracranial pressure parameters in the epidural space and in brain parenchyma in patients with hydrocephalus
Authors:
- Per Kristian EidePEPer Kristian Eide
- Wilhelm SortebergWSWilhelm Sorteberg
- 1 December 2010
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 113 (6) , 1317-1325
Abstract
Object: In this study, the authors compare simultaneous measurements of static and pulsatile pressure parameters in the epidural space and brain parenchyma of hydrocephalic patients. Methods: Simultaneous intracranial pressure (ICP) signals from the epidural space (ICPEPI) and the brain parenchyma (ICPPAR) were compared in 12 patients undergoing continuous ICP monitoring as part of their diagnostic workup for hydrocephalus. The static ICP was characterized by mean ICP and the frequency of B waves quantified in the time domain, while the pulsatile ICP was determined from the cardiac beat–induced single ICP waves and expressed by the ICP pulse pressure amplitude (dP) and latency (dT; that is, rise time). Results: The 12 patients underwent a median of 22.5 hours (range 5.9–24.8 hours) of ICP monitoring. Considering the total recording period of each patient, the mean ICP (static ICP) differed between the 2 compartments by ≥ 5 mm Hg in 8 patients (67%) and by ≥ 10 mm Hg in 4 patients (33%). In contrast, for every patient the ICP pulse pressure readings from the 2 compartments showed near-identical results. Consequently, when sorting patients to shunt/no shunt treatment according to pulsatile ICP values, selection was independent of sensor placement. The frequency of B waves also compared well between the 2 compartments. Conclusions: The pulsatile ICP is measured with equal confidence from the ICPEPI and ICPPAR signals. When using the pulsatile ICP for evaluation of hydrocephalic patients, valid measurements may thus be obtained from pressure monitoring in the epidural space. Recorded differences in the mean ICP between the epidural space and the brain parenchyma are best explained by differences in the zero setting of different sensors.Keywords
This publication has 39 references indexed in Scilit:
- Diagnostic Intracranial Pressure Monitoring and Surgical Management in Idiopathic Normal Pressure HydrocephalusNeurosurgery, 2010
- Intracranial pressure monitoring with the Neurodur-P epidural sensor: a prospective study in patients with adult hydrocephalus or idiopathic intracranial hypertensionJournal of Neurosurgery, 2008
- Continuous intraventricular pressure monitoring for diagnosis of normal-pressure hydrocephalusActa Neurochirurgica, 2007
- Intracranial Pressure during Wakefulness and Sleep in 55 Adult Patients with Chronic HydrocephalusNeurosurgery, 2006
- Monitoring and interpretation of intracranial pressureJournal of Neurology, Neurosurgery & Psychiatry, 2004
- Reliability of Epidural Pressure Measurement in Clinical Practice: Behavior of Three Modern Sensors during Simultaneous Ipsilateral Intraventricular or Intraparenchymal Pressure MeasurementNeurosurgery, 1998
- A comparison of extradural and intraparenchymatous intracranial pressures in head injured patientsIntensive Care Medicine, 1995
- Pulse amplitude and volume-pressure relationships in experimental hydrocephalusActa Neurochirurgica, 1989
- Simultaneous Recording of the Ventricular Fluid Pressure and the Epidural PressureEuropean Neurology, 1972
- MINIATURE TRANSDUCER FOR INTRACRANIAL PRESSURE MONITORING IN MANActa Neurologica Scandinavica, 1970