Moving-Table MR Angiography of the Peripheral Runoff Vessels: Comparison of Body Coil and Dedicated Phased Array Coil Systems
Authors:
- 1 May 2003
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 180 (5) , 1365-1373
Abstract
OBJECTIVE. The aim of our study was to compare the signal-to-noise ratio and the diagnostic accuracy of moving-table MR angiography of the peripheral arteries with body coil and dedicated phased array coil systems. SUBJECTS AND METHODS. Forty patients were examined with digital subtraction angiography and moving-table MR angiography with a 1.5-T MR imaging system either with a body coil (n = 20) or with a dedicated phased array coil (n = 20). The timing of contrast material was performed with real-time MR fluoroscopy. RESULTS. For the iliac artery, upper leg, and lower leg, the mean values for signal-to-noise ratios were 56, 51, and 17, respectively, for the body coil, and 54, 74, and 64, respectively, for the dedicated phased array coil. For the body coil, sensitivity and specificity in identifying stenosis greater than 50% and occlusions were 100% and 96%, respectively, for the iliac arteries, and 100% and 96%, respectively, for the upper leg. For the dedicated phased array coil, sensitivity and specificity for stenosis greater than 50% and occlusions were 100% and 96%, respectively, for the iliac arteries, and 100% and 98%, respectively, for the upper leg. Sensitivity and specificity were inferior for the body coil (88% and 85%) compared with the dedicated phased array coil (100% and 96%) in the lower leg. A significant difference of the mean values of contrast-to-noise ratio was found before and after subtraction for the dedicated phased array coil and body-coil techniques (Student's t test, p < 0.01). CONCLUSION. In comparison with the body coil, the dedicated peripheral phased array surface coil system improves signal-to-noise ratio for the upper and lower leg and diagnostic accuracy in the lower leg.Keywords
This publication has 21 references indexed in Scilit:
- Dynamic Contrast-Enhanced MR Angiography from the Distal Aorta to the Ankle Joint with a Step-by-Step TechniqueAmerican Journal of Roentgenology, 2000
- Contrast-Enhanced Subtraction MR Angiography in Occlusive Disease of the Pelvic and Lower Limb Arteries: Results of a Prospective Intraindividual Comparative Study with Digital Subtraction Angiography in 76 PatientsJournal of Computer Assisted Tomography, 1999
- Aortoiliac and Lower Extremity Arteries: Comparison of Three-dimensional Dynamic Contrast-enhanced Subtraction MR Angiography and Conventional AngiographyRadiology, 1999
- Gadolinium-enhanced three-dimensional MR angiography of the entire aorta and iliac arteries with dynamic manual table translation.Radiology, 1998
- Three-dimensional high-resolution dynamic contrast-enhanced MR angiography of the pelvis and lower extremities with use of a phased array coil and subtraction: diagnostic accuracy.Journal of Magnetic Resonance Imaging, 1998
- Gadolinium-enhanced three-dimensional MR angiography of the aorta and peripheral arteries: evaluation of a multistation examination using two gadopentetate dimeglumine infusions.American Journal of Roentgenology, 1998
- Contrast‐enhanced 3D MRA with centric ordering in k space: A preliminary clinical experience in imaging the abdominal aorta and renal and peripheral arterial vasculatureJournal of Magnetic Resonance Imaging, 1998
- Arteriographic complications in the DSA era.Radiology, 1992
- Adverse reactions to ionic and nonionic contrast media. A report from the Japanese Committee on the Safety of Contrast Media.Radiology, 1990
- Contrast Material-Induced Renal Failure in Patients with Diabetes Mellitus, Renal Insufficiency, or BothNew England Journal of Medicine, 1989