Initial Clinical Trial of Robot of Endovascular Treatment with Force Feedback and Cooperating of Catheter and Guidewire
Open Access
- 16 April 2018
- journal article
- research article
- Published by Wiley in Applied Bionics and Biomechanics
- Vol. 2018, 1-10
Abstract
To evaluate the feasibility and safety of the robot of endovascular treatment (RobEnt) in clinical practice, we carried out a cerebral angiography using this robot system. We evaluated the performance of application of the robot system to clinical practice through using this robotic system to perform the digital subtraction angiography for a patient who was suspected of suffering intracranial aneurysm. At the same time, through comparing the postoperative head nuclear magnetic and blood routine with the preoperative examination, we evaluated the safety of application of the robot system to clinical practice. We performed the robot system to complete the bilateral carotid artery and bilateral vertebral arteriography. The results indicate that there was no obvious abnormality in the patient’s cerebral artery. No obvious abnormality was observed in the examination of patients’ check-up, head nuclear magnetism, and blood routine after the digital subtraction angiography. From this clinical trial, it can be observed that the robot system can perform the operation of cerebral angiography. The robot system can basically complete the related observation indexes, and its accuracy, effectiveness, stability, and safety basically meet the requirements of clinical application in neurointerventional surgery.Keywords
Funding Information
- Scientific and Technological Projects of Science and Technology Commission of Beijing (Z161100002616002, 2015AA043202)
- National High Technology Research and Development Program (Z161100002616002, 2015AA043202)
This publication has 16 references indexed in Scilit:
- Begin Again and Continue With Life: A Qualitative Study on the Experiences of Cardiac Rehabilitation PatientsJournal of Nursing Research, 2017
- AORN Ergonomic Tool 4: Solutions for Prolonged Standing in Perioperative SettingsAORN Journal, 2011
- [Epidemiologic characteristics of cerebrovascular disease mortality in China, 2004-2005].2010
- Remote Robotic Navigation and Electroanatomical Mapping for Ablation of Atrial FibrillationCirculation: Arrhythmia and Electrophysiology, 2009
- Occupational Health Hazards in the Interventional Laboratory: Time for a Safer EnvironmentJournal of Vascular and Interventional Radiology, 2008
- Remote navigation systems in electrophysiologyEP Europace, 2008
- OCCUPATIONAL RADIATION DOSES TO OPERATORS PERFORMING CARDIAC CATHETERIZATION PROCEDURESHealth Physics, 2008
- Experience of robotic catheter ablation in humans using a novel remotely steerable catheter sheathJournal of Interventional Cardiac Electrophysiology, 2008
- View-Synchronized Robotic Image-Guided Therapy for Atrial Fibrillation AblationCirculation, 2007
- Robotic Magnetic Navigation for Atrial Fibrillation AblationJACC, 2006