Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry
Authors:
- Josef NovotnyJNJosef Novotny
- Jagdish P. BhatnagarJBJagdish P. Bhatnagar
- Mubina A. QuaderMQMubina A. Quader
- Greg BednarzGBGreg Bednarz
- L. Dade LunsfordLLL. Dade Lunsford
- M. Saiful HuqMHM. Saiful Huq
- 20 April 2009
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 36 (5) , 1768-1774
Abstract
Three types of films, Kodak EDR2, Gafchromic EBT, and Gafchromic MD‐V2‐55, were used to measure relative output factors of 4 and 8 mm collimators of the Leksell Gamma Knife Perfexion. The optical density to dose calibration curve for each of the film types was obtained by exposing the films to a range of known doses. Ten data points were acquired for each of the calibration curves in the dose ranges from 0 to 4 Gy, 0 to 8 Gy, and 0 to 80 Gy for Kodak EDR2, Gafchromic EBT, and Gafchromic MD‐V2‐55 films, respectively. For the measurement of relative output factors, five films of each film type were exposed to a known dose. All films were scanned using EPSON EXPRESSION 10000 XL scanner with 200 dpi resolution in 16 bit gray scale for EDR2 film and 48 bit color scale for Gafchromic films. The scanned images were imported in the red channel for both Gafchromic films. The background corrections from an unexposed film were applied to all films. The output factors obtained from film measurements were in a close agreement both with the Monte Carlo calculated values of 0.924 and 0.805 for 8 and 4 mm collimators, respectively. These values are provided by the vendor and used as default values in the vendor's treatment planning system. The largest differences were noted for the Kodak EDR 2 films (−2.1% and −4.5% for 8 and 4 mm collimators, respectively). The best agreement observed was for EBT Gafchromic film (−0.8% and differences for 8 and 4 mm collimators, respectively). Based on the present values, no changes in the default relative output factor values were made in the treatment planning system.Keywords
This publication has 33 references indexed in Scilit:
- Dosimetric comparison of the Leksell Gamma Knife Perfexion and 4CJournal of Neurosurgery, 2008
- THE LEKSELL GAMMA KNIFE PERFEXION AND COMPARISONS WITH ITS PREDECESSORSOperative Neurosurgery, 2007
- A comparison of the gamma knife model C and the Automatic Positioning System with Leksell model BJournal of Neurosurgery, 2005
- The Leksell Gamma Knife Model U versus Model C: A Quantitative Comparison of Radiosurgical Treatment ParametersNeurosurgery, 2004
- Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beamsPhysics in Medicine & Biology, 2002
- Precision dosimetry for narrow photon beams used in radiosurgery—Determination of Gamma Knife® output factorsMedical Physics, 2002
- An Evaluation of the Model C Gamma Knife with Automatic Patient PositioningNeurosurgery, 2002
- First clinical experience with the automatic positioning system and Leksell gamma knife Model CJournal of Neurosurgery, 2000
- Physics of gamma knife approach on convergent beams in stereotactic radiosurgeryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990
- Stereotactic radiosurgery of the brain using the first United States 201 cobalt-60 source gamma knifeNeurosurgery, 1989