Single-Particle Fluorescence Spectrometer for Ambient Aerosols
Authors:
Open Access
- 1 August 2003
- journal article
- research article
- Published by Taylor & Francis in Aerosol Science and Technology
- Vol. 37 (8) , 628-639
Abstract
A fluorescence particle spectrometer (FPS) for real-time measurement of the fluorescence spectra of aerosol particles in the size range 1-10 w m diameter is reported. The prototype FPS has a sufficiently high sample rate (from 5 to 28 l/min for 3.5 w m to 11 w m diameter particles) to measure aerosol within buildings at practical rates (from 1 up to 600 particle fluorescence spectra per minute). Previously reported bioaerosol prototype detectors for measurement of single particle spectra (Pan et al., Opt. Lett ., 24, 116-118 (1999); Hill et al., Field Anal. Chem. Tech ., 3, 221-239 (1999)) were unable to sample the ambient environment; air containing particles had to be forced under pressure into a sample cell. In addition, sample rates were so small (less than 0.01 l/min) as to be impractical for most applications. The present design overcomes these deficiencies by the use of an airtight cell that highly concentrates micrometer-sized particles. A virtual impactor first concentrates aerosol particles, which are then drawn under negative pressure through an aerodynamic focusing nozzle in the inlet of the instrument, through the sample region, providing further concentration. The rate of particle spectra measured by the FPS increases significantly when the particle inlet is within a few meters of some common sources of indoor biological particles, e.g., a person coughing, sneezing, or rubbing his skin, or the presence of a dog. The spectra obtained have a variety of spectral shapes. The FPS may be useful in a variety of areas, e.g., in studying and monitoring airborne particles that cause diseases or allergies.Keywords
This publication has 45 references indexed in Scilit:
- Diurnal Distribution of Total and Culturable Atmospheric Bacteria at a Rural SiteAerosol Science and Technology, 1999
- Detection of Bioaerosols Using Multiwavelength UV Fluorescence SpectroscopyAerosol Science and Technology, 1999
- Aerosol Fluorescence Spectrum Analyzer for Rapid Measurement of Single Micrometer-Sized Airborne Biological ParticlesAerosol Science and Technology, 1998
- Light Scattering Instrumentation for Aerosol Studies: An Historical OverviewAerosol Science and Technology, 1997
- Conditional-firing aerosol-fluorescence spectrum analyzer for individual airborne particles with pulsed 266-nm laser excitationOptics Letters, 1996
- Mass monitoring of carbonaceous aerosols with a spectrophoneAtmospheric Environment (1967), 1979
- Experimental studies of optical particle countersAtmospheric Environment (1967), 1974
- Detection and Sizing of Small Particles in an Open Cavity Gas LaserApplied Optics, 1972
- Angular truncation error in the integrating nephelometerAtmospheric Environment (1967), 1970
- A stabilized integrating nephelometer for visibility studiesAtmospheric Environment (1967), 1970