On-X mitral valve leaflet embolization presenting as acute pulmonary oedema
Authors:
Open Access
- 4 May 2022
- journal article
- case report
- Published by Springer Nature in The Egyptian Heart Journal
- Vol. 74 (1) , 1-4
Abstract
Background: Structural valve failure in the form of leaflet fracture and embolization is an uncommon complication. Severe valve regurgitation with pulmonary oedema in a patient with a history of a prosthetic cardiac valve requires urgent diagnosis. An echocardiogram is essential in identifying the cause of valve failure. Case presentation: We report a case of a 21-year-old male patient who presented with acute pulmonary oedema due to fracture and embolization of a leaflet of an On-X mitral valve four years post-implant. He underwent an emergency redo mitral valve replacement with a good outcome. Conclusions: Structural valve failure with leaflet fracture and embolization is an extremely rare complication and may be fatal. Good clinical acumen is required for prompt diagnosis in any patient who presents with a circulatory collapse in the presence of a history of previous cardiac surgery. Once the diagnosis is made, the patient should undergo emergency redo valve replacement surgery.Keywords
This publication has 6 references indexed in Scilit:
- Cardiogenic Shock Due to Leaflet Migration of On-X Aortic Mechanical ProsthesisThe Annals of Thoracic Surgery, 2020
- Survival After Mitral Valve Replacement for Leaflet Escape in a Contemporary On-X Mechanical ValveThe Annals of Thoracic Surgery, 2019
- Spontaneous leaflet fracture resulting in embolization from mechanical valve prosthesesJournal of Cardiac Surgery, 2018
- Sudden haemodynamic collapse caused by leaflet escape of the contemporary On-X mechanical valveEuropean Journal of Cardio-Thoracic Surgery, 2018
- Acute mitral valve dysfunction due to leaflet escape in a Tri-technologies bileaflet mechanical valveEuropean Heart Journal - Cardiovascular Imaging, 2007
- Leaflet Escape in a New Bileaflet Mechanical ValveCirculation, 2003