Abstract
A spectrofluorimetric method is proposed for the determination of human serum albumin (HSA) and bovine serum albumin (BSA) using terbium-danofloxacin (Tb3+-Dano) as a fluorescent probe. These proteins remarkably enhance the fluorescence intensity of the Tb3+-Dano complex at 545nm, and the enhanced fluorescence intensity of Tb3+-Dano is proportional to the concentration of proteins (HSA and BSA). Optimum conditions for the determination of HSA were investigated and found that the maximum response was observed at: p H = 7 . 8 , [Tb3+] = 8 . 5 × 1 0 − 5 mol L−1, [Dano] = 1 . 5 × 1 0 − 4 mol L−1. The calibration graphs for standard solutions of BSA, HSA, and plasma samples of HSA were linear in the range of 0 . 2 × 1 0 − 6 − 1 . 3 × 1 0 − 6 mol L−1, 0 . 2 × 1 0 − 6 − 1 . 4 × 1 0 − 6 mol L−1, and 0 . 2 × 1 0 − 6 − 1 × 1 0 − 6 mol L−1, respectively. The detection limits (S/N = 3) for BSA, HSA, and plasma sample of HSA were 8 . 7 × 1 0 − 8 mol L−1, 6 . 2 × 1 0 − 8 mol L−1, and 8 . 1 × 1 0 − 8 mol L−1, respectively. The applicability of the method was checked using a number of real biological plasma samples and was compared with the UV spectrometric reference method. The results was showed that the method could be regarded as a simple, practical, and sensitive alternative method for determination of albumin in biological samples.

This publication has 28 references indexed in Scilit: