MR temperature measurement in liver tissue at 0.23 T with a steady-state free precession sequence
Authors:
Open Access
- 22 April 2002
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 47 (5) , 940-947
Abstract
MRI can be used for monitoring temperature during a thermocoagulation treatment of tumors. The aim of this study was to demonstrate the suitability of a 3D steady‐state free precession sequence (3D Fast Imaging with Steady‐State Precession, 3D TrueFISP) for MR temperature measurement at 0.23 T, and to compare it to the spin‐echo (SE) and spoiled 3D gradient‐echo (3D GRE) sequences. The optimal flip angle for the TrueFISP sequence was calculated for the best temperature sensitivity in the image signal from liver tissue, and verified from the images acquired during the thermocoagulation of excised pig liver. Factors influencing the accuracy of the measured temperatures are discussed. The TrueFISP results are compared to the calculated values of optimized SE and 3D GRE sequences. The accuracy of TrueFISP in the liver at 0.23 T, in imaging conditions used during thermocoagulation procedures, is estimated to be ±3.3°C for a voxel of 2.5 × 2.5 × 6 mm3 and acquisition time of 18 s. For the SE and GRE sequences, with similar resolution and somewhat longer imaging time, the uncertainty in the temperature is estimated to be larger by a factor of 2 and 1.2, respectively. Magn Reson Med 47:940–947, 2002.Keywords
This publication has 22 references indexed in Scilit:
- Interventional MRI of the breast: minimally invasive therapyEuropean Radiology, 2000
- Time and temperature dependence of MR parameters during thermal coagulation of ex vivo rabbit muscleMagnetic Resonance in Medicine, 1998
- Percutaneous Radiofrequency Tissue Ablation: Does Perfusion-mediated Tissue Cooling Limit Coagulation Necrosis?Journal of Vascular and Interventional Radiology, 1998
- Noninvasive MRI thermometry with the proton resonance frequency method: Study of susceptibility effectsMagnetic Resonance in Medicine, 1995
- Measurements of the effect of storage at various temperatures on the T1 of ex vivo tissuesMedical Physics, 1995
- Temperature dependence of canine brain tissue diffusion coefficient measuredin vivowith magnetic resonance echo-planar imagingInternational Journal of Hyperthermia, 1995
- Modeling and observation of temperature changes in Vivo using MRIMagnetic Resonance in Medicine, 1994
- MR temperature mapping of focused ultrasound surgeryMagnetic Resonance in Medicine, 1994
- Tissue temperature monitoring for thermal interventional therapy: Comparison of T1‐weighted MR sequencesJournal of Magnetic Resonance Imaging, 1994
- Motion‐insensitive, steady‐state free precession imagingMagnetic Resonance in Medicine, 1990