Stimulation Characteristics of a Steroid‐Eluting Electrode Compared with Three Conventional Electrodes
Authors:
- 1 February 1990
- journal article
- editorial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 13 (2) , 134-137
Abstract
Stimulation thresholds of a steroid‐eluting electrode (Medtronic CapSure 4003, 9 patients) were compared with three conventional electrodes (carbon‐tipped lead, Siemens‐Elema 423 S/60, 10 patients; Elgiloy‐tipped lead, Cordis Encor, 10 patients; platinum‐tipped lead, Telectronix Laserdish 030–276, 9 patients). Voltage thresholds were determined during implantation, 1–3 days, 6 weeks, and 6 months postimplantation. No significant difference among the four electrodes was found in regard to stimulation and sensing behavior during implantation. Compared to intraoperative measurements, mean increase in voltage threshold and absolute voltage thresholds were substantially Jess for the steroid‐eluting electrode than for the other tested electrodes 6 weeks and 6 months after implantation. The rheobasechronaxie product, a parameter of the stimulation performance of electrodes, underlined the superior pacing characteristics of the steroid‐eluting electrode. To achieve low chronic stimulation threshold by pharmacological means is an attractive direction for future electrode technology.Keywords
This publication has 12 references indexed in Scilit:
- Cellular electrophysiologic effects of dexamethasone sodium phosphate: implications for cardiac stimulation with steroid-eluting electrodesInternational Journal of Cardiology, 1989
- Paired Comparisons of Steroid‐Eluting and Nonsteroid Endocardial Pacemaker Leads in Dogs: Electrical Performance and Morphologic AlterationsPacing and Clinical Electrophysiology, 1988
- The Porous Titanium Steroid Eluting Electrode: A Double Blind Study Assessing the Stimulation Threshold Effects of SteroidPacing and Clinical Electrophysiology, 1988
- Long‐Term Performance of Endocardial Leads with Steroid‐Eluting ElectrodesPacing and Clinical Electrophysiology, 1986
- Comparison of Vitreous Carbon and Elgiloy Transvenous Ventricular Pacing LeadsPacing and Clinical Electrophysiology, 1985
- Erfahrungen mit telemetrie-gestützten Schrittmacherkontrollen bei Patienten mit VVI-SchrittmachernDMW - Deutsche Medizinische Wochenschrift, 1985
- Schrittmachertherapie mit steroidhaltiger Elektrode: Bessere Stimulations- und Detektionsbedingungen durch Steroid-Freisetzung aus der SondenspitzeDMW - Deutsche Medizinische Wochenschrift, 1984
- Comparison of Pacing Electrodes of Different Shape and Material—RecommendationsPacing and Clinical Electrophysiology, 1983
- Electrode‐Heart Interface: Definition of the Ideal ElectrodePacing and Clinical Electrophysiology, 1983
- Stimulation and sensing thresholds for cardiac pacing: Electrophysiologic and technical aspectsProgress in Cardiovascular Diseases, 1981