Sample transport and/or storage can cause falsely low HbA1c levels in blood cells measured by enzymatic assay
Authors:
- 8 November 2019
- journal article
- research article
- Published by Springer Nature in Diabetology International
- Vol. 11 (2) , 155-157
Abstract
Although HbA1c measurement by enzymatic assay (EA-HbA1c) is widely used in health-screening settings in Japan, recent studies have suggested lower EA-HbA1c levels as compared with HbA1c levels measured by high-performance liquid chromatography (HPLC-HbA1c). Hypothesizing that falsely low levels of EA-HbA1c are attributable to hemolysis caused by sample transport and/or storage, we measured EA-HbA1c in blood cells and whole blood after sample transport and compared them with HPLC-HbA1c levels. Blood samples were collected from ten non-diabetic individuals into sodium fluoride-containing blood collection tubes and immediately measured for EA-HbA1c in blood cells. After transport, the blood samples were again subjected to measurement of EA-HbA1c levels in blood cells and whole blood the following day. These EA-HbA1c levels were compared with HPLC-HbA1c levels. EA-HbA1c levels in blood cells measured immediately after sample collection did not significantly differ from HPLC-HbA1c levels. Transported blood samples showed hemolysis and significantly lower EA-HbA1c levels in blood cells, as compared with HPLC-HbA1c levels, whereas no significant difference was observed between EA-HbA1c levels in whole blood and HPLC-HbA1c levels. Transported blood samples showed hemolysis and falsely low EA-HbA1c levels in blood cells. Hemolysis caused by sample transport and/or storage might be responsible for the falsely low EA-HbA1c levels. This should be kept in mind, because falsely low HbA1c levels may lead to a false-negative diagnosis of diabetes.Keywords
This publication has 5 references indexed in Scilit:
- HbA1c levels measured by enzymatic assay during off-site health checkups are lower than those measured by on-site HPLC assayDiabetology International, 2019
- Evaluating new HbA1c methods for adoption by the IFCC and NGSP reference networks using international quality targetsClinical Chemistry and Laboratory Medicine (CCLM), 2017
- Haemoglobin variants may cause significant differences in haemoglobin A1c as measured by high-performance liquid chromatography and enzymatic methods in diabetic patients: a cross-sectional studyAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2016
- Use of Fructosyl Peptide Oxidase for HbA1c AssayJournal of Diabetes Science and Technology, 2015
- Glycated hemoglobin in fractionated erythrocytes.Clinical Chemistry, 1989