Measurement of local strains induced into the femur by trochanteric Gamma nail implants with one or two distal screws
- 28 February 2006
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 29 (1) , 38-47
Abstract
This study aims to evaluate experimentally the behaviour after fracture consolidation of two intramedullary Gamma nail implants, having one (G1) or two (G2) distal screws, respectively. Nails have been implanted into standardized synthetic femora, instrumented with strain gauges. Strains measurements, supported by finite element numerical modelling, showed that the G2 implant, although ensuring higher flexional and torsional stiffness, can lead to localized contacts that occur between the tip of the nail and the femoral endosteum. This might be one of the reasons of the complications associated with pain in the mid-portion of the thigh after implantation which has been reported in several clinical studies when Gamma nails with two distal screws are used.Keywords
This publication has 25 references indexed in Scilit:
- Intramedullary nails with two lag screwsClinical Biomechanics, 2004
- Mechanical simulation of muscle loading on the proximal femur: analysis of cemented femoral component migration with and without muscle loadingClinical Biomechanics, 2003
- Intramedullary nails: some design features of the distal endMedical Engineering & Physics, 2003
- Necrosis of the femoral head after fixation of trochanteric fractures with Gamma Locking NailInjury, 2003
- Finite element study of trochanteric gamma nail for trochanteric fractureMedical Engineering & Physics, 2003
- Biomechanical Evaluation of the Proximal Femoral NailClinical Orthopaedics and Related Research, 2002
- Mechanical validation of whole bone composite femur modelsJournal of Biomechanics, 1996
- Comparison of dynamic hip screw and gamma nail: a prospective, randomized, controlled trialInjury, 1995
- A review of biomechanical studies of intramedullary nailsMedical Engineering & Physics, 1995
- Failure of the Gamma Nail in a Highly Unstable Proximal Femur Fracture: Report of Four Cases Encountered in The NetherlandsJournal of Orthopaedic Trauma, 1995