Comparison of Minimally Invasive Inductively Coupled Plasma–Mass Spectrometry Approaches for Strontium Isotopic Analysis of Medieval Stained Glass with Elevated Rubidium and Rare-Earth Element Concentrations
Open Access
- 9 July 2021
- journal article
- research article
- Published by American Chemical Society (ACS) in ACS Omega
- Vol. 6 (28) , 18110-18122
Abstract
Different approaches for the determination of the 87Sr/86Sr isotope ratio of high-Rb glass are compared in this work to assess the suitability of minimally invasive approaches for applications on medieval stained glass (from the ancient Abbey of Stavelot in Belgium). It was found that pneumatic nebulization multicollector inductively coupled plasma–mass spectrometry (PN-MC-ICP-MS) after acid digestion and chromatographic isolation of the target analyte out of the sample matrix can still be seen as the preferred method for the high-precision isotopic analysis of Sr in glass with high Rb and rare-earth element (REE) concentrations. Alternatively, the use of laser ablation (LA) for sample introduction is a powerful technique for the direct analysis of solid samples. However, both the high Rb/Sr ratios in the samples of interest and the presence of REEs at sufficiently high concentrations lead to a large bias in LA-MC-ICP-MS, which cannot be corrected for, even by operating the MC-ICP-MS instrument at higher mass resolution and/or using mathematical corrections. It was demonstrated that LA tandem-ICP-MS (LA-ICP-MS/MS) using CH3F/He as the reaction gas to overcome spectral overlap in a mass-shift approach (chemical resolution) provides a viable alternative when (quasi) nondestructive analysis is required. This approach relies on the monitoring of Sr+ (m/z = 86, 87, and 88) ions as the corresponding SrF+ reaction product ions (m/z = 105, 106, and 107), thus avoiding the occurrence of spectral interference. Self-evidently, the isotope ratio precision attainable using sequential quadrupole-based ICP-MS instrumentation (0.3% RSD) was found to be significantly worse than that of high-precision MC-ICP-MS (0.03% RSD) with simultaneous detection, although it was still fit for the purpose of current applications. In addition to Sr isotopic analysis, the REE patterns and their potential influence on the Sr isotopic composition were evaluated by LA-ICP-MS.Keywords
Funding Information
- Fonds Wetenschappelijk Onderzoek (12ZA320N, G.0C43.15)
This publication has 41 references indexed in Scilit:
- Assessment of Nanosecond Laser Ablation Multi‐Collector Inductively Coupled Plasma‐Mass Spectrometry for Pb and Sr Isotopic Determination in Archaeological Glass: Mass Bias Correction Strategies and Results for Corning Glass Reference MaterialsGeostandards and Geoanalytical Research, 2017
- Recirculation of biomass ashes onto forest soils: ash composition, mineralogy and leaching propertiesWaste Management, 2017
- Laser ablation-tandem ICP-mass spectrometry (LA-ICP-MS/MS) for direct Sr isotopic analysis of solid samples with high Rb/Sr ratiosJournal of Analytical Atomic Spectrometry, 2015
- Tandem ICP-mass spectrometry for Sr isotopic analysis without prior Rb/Sr separationJournal of Analytical Atomic Spectrometry, 2015
- Isotope analysis of English forest glass from the Weald and StaffordshireJournal of Analytical Atomic Spectrometry, 2012
- Potash – a key raw material of glass batch for Bohemian glasses from 14th–17th centuries?Journal of Archaeological Science, 2012
- The chemical composition including the Rare Earth Elements of the three major glass types of Europe and the Orient used in late antiquity and the Middle AgesGeochemistry, 2011
- The roots of provenance: glass, plants and isotopes in the Islamic Middle East2009
- 87Sr/86Sr sourcing of ponderosa pine used in Anasazi great house construction at Chaco Canyon, New MexicoJournal of Archaeological Science, 2005
- Potassium–Calcium Glass: New Data and Experiments*Archaeometry, 2004