Performance evaluation of hematology analyzer XN-10 (Blood Bank mode) and its comparison with flow-cytometer for quality control (QC) of leukodepleted blood components: Pilot Study
Open Access
- 22 July 2025
- journal article
- Published by Wolters Kluwer Health in Asian Journal of Transfusion Science
- Vol. 20 (1) , 13-21
Abstract
BACKGROUND: The quality control of leukodepleted blood components is based on residual white blood cells (rWBCs). Routinely rWBC in leukodepleted are performed flow cytometer (FCM). Newer hematology analyzers like Sysmex XN series can count rWBC accurately. STUDY DESIGN AND MATERIALS AND METHODS: There were two study objectives: (1) to evaluate analytical performance of XN blood bank (BB) mode, and (2) to compare the measurement of rWBC and residual red blood cell between the XN BB mode and flow cytometer (FCM), considered “gold standard.” This cross sectional analytical study was conducted on three type of blood components; leukodepleted random donor platelets (LRDP), leukodepleted single donor platelets (LSDP), and leukodepleted red blood cells (LRBC) to determine performance analysis; precision (coefficients of variation), sample stability, and carry over of XN BB. Correlation and linearity was compared between XN BB and FCM. RESULTS: In precision analysis, LRBC, LRDP, and LSDP showed good repeatability (P < 0.0001). Sample stability results revealed stability of LRDP and LRBC samples till 48 h, and LSDP sample was least stable as rWBC counts changed after 5 h. No carry over was observed in any blood component in the study. While comparing XN BB and FCM, a strong correlation was observed between the two, with P > 0.05. In addition, there was an excellent linearity between expected and observed WBC counts for LRBC and LSDP in BB mode (r2 = 0.88). DISCUSSION: The analyzer performed very well with respect to precision, carry over, and stability. The new BB mode of the XN 10 hematology analyzer also showed highly comparable results with FCM, in terms of correlation and linearity, to determine rWBC. CONCLUSION: The study concluded the use of BB mode as an alternative for flow cytometric analysis for the determination of rWBC in erythrocyte concentrates, platelet concentrate, and plasma.Keywords
This publication has 9 references indexed in Scilit:
- Residual red cells in blood components: A multisite study of fully automated enumeration using a hematology analyzerTransfusion, 2020
- Improved accuracy in counting residual white blood cells in red cell concentrates using new blood bank mode software of SYSMEX XN‐1000 hematology analyzerTransfusion, 2020
- The use of a hematology analyzer with a new generation of software as an alternative to flow cytometry for enumerating residual white blood cells in blood componentsTransfusion, 2019
- Enumeration of residual white blood cells in leukoreduced blood products: Comparing flow cytometry with a portable microscopic cell counterTransfusion and Apheresis Science, 2015
- ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte countingInternational Journal of Laboratory Hematology, 2014
- Stability: Recommendation for Best Practices and Harmonization from the Global Bioanalysis Consortium Harmonization TeamThe AAPS Journal, 2014
- Leucodepletion and Blood ProductsMedical Journal Armed Forces India, 2006
- Guidelines on the clinical use of leucocyte-depleted blood components. British Committee for Standards in Haematology, Blood Transfusion Task Force.1998
- Guidelines on the clinical use of leucocyte‐depleted blood componentsTransfusion Medicine, 1998