The effect of the subcutaneous fat on the transfer of current through skin and into muscle
- 8 April 2008
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 30 (9) , 1168-1176
Abstract
The present investigation was conducted to see the effect of subcutaneous fat on the transmission characteristics of an electrical stimulus applied to the skin and conducted to skeletal muscle. Two groups of subjects participated. In one, the subjects were three males and three females whose average age was 24.6+/-1.5 years, average weight was 74.8+/-18.2kg, and average height was 176.4+/-10.3cm. The other was a group of 30 subjects who average age was 26.2+/-1.9 years, average height 177.3+/-11.5cm, average weight 92.4+/-19.8kg. Electrical stimulation was applied above the quadriceps muscle at a current of 5mA and with sine and square wave stimulation at a frequency of 30Hz and a pulse width of 250micros. Current movement was measured on the skin and into muscle with surface and needle electrodes. The results showed that the thickness of the subcutaneous fat layer was directly related to signal loss from the skin (correlation between subcutaneous fat thickness and RC time constant was 0.96, p<0.001). Because of the subcutaneous fat layer and the resulting capacitance, an RC low pass filter is created such that square wave stimuli are not transmitted well into muscle whereas sine wave stimuli pass easily. Thus, when considering surface stimulation of nerve or muscle, any volume conductor model must take subcutaneous fat into consideration since the RC low pass filter created by fat will filter surface signals or, conversely, signals such as EMG which are generated in muscle but measured on the surface of the skin.Keywords
This publication has 25 references indexed in Scilit:
- A re-evaluation of modelling of the current flow between electrodes: consideration of blood flow and woundsJournal of Medical Engineering & Technology, 2007
- Current distribution under electrodes in relation to stimulation current and skin blood flow: are modern electrodes really providing the current distribution during stimulation we believe they are?Journal of Medical Engineering & Technology, 2006
- Body composition modeling in the calf using an equivalent circuit model of multi-frequency bioimpedance analysisPhysiological Measurement, 2005
- Basic study of new diagnostic modality according to non-invasive measurement of the electrical conductivity of tissuesThe Journal of Medical Investigation, 2004
- A Comparison of Air Displacement Plethysmography with Three Other Techniques to Determine Body Fat in Healthy AdultsJournal of Parenteral and Enteral Nutrition, 1999
- Neuromuscular Electrical Stimulation and Dynamic Bracing as a Treatment for Upper-Extremity Spasticity in Children with Cerebral PalsyJournal of Hand Surgery (European Volume), 1999
- Volume conduction models for surface EMG; confrontation with measurementsJournal of Electromyography and Kinesiology, 1997
- A parallel broadband real-time system for electrical impedance tomographyPhysiological Measurement, 1996
- A model study of extracellular stimulation of cardiac cellsIEEE Transactions on Biomedical Engineering, 1993
- Comparison of isometric muscle training and electrical stimulation supplementing isometric muscle training in the recovery after major knee ligament surgeryThe American Journal of Sports Medicine, 1979