Evaluation of Contact Resistance Between Catheters and Guidewires in Pancreatobiliary Endoscopy: Influence of Material and Mechanical Properties

Authors:

Abstract
Background: Guidewire manipulation is critical in pancreatobiliary endoscopy and is always performed in combination with catheters. However, the impact of catheter–guidewire interactions on contact resistance remains unclear. This study aimed to evaluate the contact resistance between various commercially available catheters and guidewires. Methods: An ex vivo experiment was conducted using 9 catheters and 15 guidewires. Contact resistance during guidewire insertion and withdrawal was measured for each catheter–guidewire combination. Subgroup analyses were performed by categorizing the catheters and guidewires according to their materials and mechanical properties. Results: In total, 750 insertion and withdrawal tests were performed. High‐density polyethylene (HDPE) catheters consistently exhibited the lowest contact resistance across all guidewires. Among the guidewire materials, perfluoroalkoxy alkanes (PFAs) showed significantly lower resistance than the other materials. Guidewires with a friction‐reducing surface had lower resistance, and stiffness showed a V‐shaped correlation with insertion resistance. Both excessively high and low stiffness increased resistance. Conclusions: Catheter–guidewire contact resistance was significantly influenced by material, surface processing, and guidewire stiffness. HDPE catheters and PFA guidewires with moderate stiffness and friction‐reducing surfaces may optimize performance. These findings offer useful insights for the selection of clinical devices and guidewire development.