Calibration of echo‐planar 2D‐selective RF excitation pulses
Open Access
- 26 October 2004
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 52 (5) , 1136-1145
Abstract
Echo‐planar radiofrequency (RF) pulses (EPP) are increasingly being used for 2D‐selective excitation in MRI. Pulse schemes of this kind are susceptible to eddy‐current effects, timing imperfections, and anisotropy of the gradient system. As a consequence, practical EPP implementations have been restricted to robust fly‐back strategies that use only every other leg of the echo‐planar trajectory for RF transmission. The present work is dedicated to enabling forward‐backward EPP with RF transmission during each k‐space segment, hence doubling the pulses' time efficiency. This is accomplished by comprehensive pulse calibration based on preparatory measurements of the system imperfections, including potential gradient anisotropy. The effectiveness of the method is demonstrated in vitro and in vivo. By doubling the speed of k‐space coverage, the proposed method enhances the potential of EPP for numerous applications. For example, motion‐sensitive techniques benefit from shorter feasible echo times (TEs) and improved excitation profiles resulting from reduced in‐pulse motion. In sequences with fast repetition, shorter EPP help reduce the overall scan duration. Alternatively, the higher time efficiency of forward‐backward EPP can enhance their spatial selectivity. Magn Reson Med 52:1136–1145, 2004.Keywords
This publication has 23 references indexed in Scilit:
- Image distortion correction in EPI: Comparison of field mapping with point spread function mappingMagnetic Resonance in Medicine, 2002
- Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imagingMagnetic Resonance in Medicine, 2002
- Assessment of Reactive Hyperaemia Using Real Time Zonal Echo-Planar Flow ImagingJournal of Cardiovascular Magnetic Resonance, 2002
- Tracking local volume 3D-echo-planar coronary artery imagingMagnetic Resonance in Medicine, 2001
- A Method to measure arbitrary k‐space trajectories for rapid MR imagingMagnetic Resonance in Medicine, 1997
- Reducing flow artifacts in echo-planar imagingMagnetic Resonance in Medicine, 1997
- Compensation of multi‐dimensional selective excitation pulses using measured k‐space trajectoriesMagnetic Resonance in Medicine, 1995
- Echo‐planar spin‐echo and inversion pulsesMagnetic Resonance in Medicine, 1993
- A k-space analysis of small-tip-angle excitationJournal of Magnetic Resonance (1969), 1989
- A method of measuring field-gradient modulation shapes. Application to high-speed NMR spectroscopic imagingJournal of Physics E: Scientific Instruments, 1987