A prospective cohort study on the short collum femoris-preserving (CFP) stem using RSA and DXA
Authors:
- Stergios LazarinisSLStergios Lazarinis
- Per MattssonPMPer Mattsson
- Jan MilbrinkJMJan Milbrink
- Hans MallminHMHans Mallmin
- Nils P HailerNHNils P Hailer
Open Access
- 23 January 2013
- journal article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 84 (1) , 32-39
Abstract
Short femoral stems have been introduced in total hip arthroplasty in order to save proximal bone stock. We hypothesized that a short stem preserves periprosthetic bone mineral density (BMD) and provides good primary stability. We carried out a prospective cohort study of 30 patients receiving the collum femoris-preserving (CFP) stem. Preoperative total hip BMD and postoperative periprosthetic BMD in Gruen zones 1-7 were investigated by dual-energy x-ray absorptiometry (DXA), stem migration was analyzed by radiostereometric analysis (RSA), and the Harris hip score (HHS) was determined. 2 patients were excluded intraoperatively and 1 patient was revised due to a deep infection, leaving 27 patients for analysis. The mean HHS increased from 49 (24-79) preoperatively to 99 (92-100) after 2 years. DXA after 1 year showed substantial loss of BMD in Gruen zone 7 (-31%), zone 6 (-19%), and zone 2 (-13%, p < 0.001) compared to baseline BMD determined immediately postoperatively. The bone loss in these regions did not recover after 2 years, whereas the more moderate bone loss in Gruen zones 1, 3, and 5 partially recovered. There was a correlation between low preoperative total hip BMD and a higher amount of bone loss in Gruen zones 2, 6 and 7. RSA showed minor micromotion of the stem: mean subsidence was 0.13 (95% CI: -0.28 to 0.01) mm and mean rotation around the longitudinal axis was 0.01º (95% CI: -0.1 to 0.39) after 2 years. We conclude that substantial loss in proximal periprosthetic BMD cannot be prevented by the use of a novel type of short, curved stem, and forces appear to be transmitted distally. However, the stems showed very small migration-a characteristic of stable uncemented implants.Keywords
This publication has 30 references indexed in Scilit:
- A prospective randomized study comparing electrochemically deposited hydroxyapatite and plasma-sprayed hydroxyapatite on titanium stemsActa Orthopaedica, 2010
- Inferior outcome after hip resurfacing arthroplasty than after conventional arthroplastyActa Orthopaedica, 2010
- A randomized study using DXA and RSA in 38 patients followed for 5 yearsActa Orthopaedica, 2010
- Mid-term results of 155 patients treated with a collum femoris preserving (CFP) short stem prosthesisInternational Orthopaedics, 2010
- Female patients with low systemic BMD are prone to bone loss in Gruen zone 7 after cementless total hip arthroplastyActa Orthopaedica, 2009
- Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stemsActa Orthopaedica, 2009
- Periprosthetic bone remodelling of a collum femoris preserving cementless titanium femoral hip replacementComputer Methods in Biomechanics and Biomedical Engineering, 2007
- Comparison of the uncemented Cone and the cemented Bimetric hip prosthesis in young patients with osteoarthritis: An RSA, clinical and radiographic studyActa Orthopaedica, 2006
- Periprosthetic proximal bone loss after uncemented hip arthroplasty is related to stem size: DXA measurements in 138 patients followed for 2–7 yearsActa Orthopaedica, 2006
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005