Dynamic Analysis of Hip Prosthesis Using Different Biocompatible Alloys
Open Access
- 1 January 2022
- journal article
- research article
- Published by ASME International in ASME Open Journal of Engineering
- Vol. 1
Abstract
In the present paper, a three-dimensional finite element model of the Charnley implant has been developed to analyze the stress–strain distribution and deformation over the stem prosthesis. Patient-specific dynamic forces have been considered for the analytical evaluation using commercial finite element code. The impact of each dynamic activity has been analyzed separately using six different biocompatible alloy materials made of titanium and cobalt-chromium. Mechanical parameters have been evaluated to envisage the longevity and functionality of the implant. The performance of different materials for each suitable gait pattern is analyzed using finite element code. Consequently, Cobalt chromium alloys (CoCrMo alloy) demonstrate better results, i.e., maximum stress, minimum deformation, and strain as compared with other materials. Every dynamic motion, viz., walking, standing up, sitting down, going upstairs, and going downstairs are found in good agreement with the safety factor for every biomaterial. Additionally, going downstairs and sitting down gait motion exhibits the maximum and minimum stress–strain level, respectively. Based on the outcome of the presurgical study, it is recommended that CoCrMo alloys should be preferred over other materials.Keywords
This publication has 34 references indexed in Scilit:
- A Dynamic Finite Element Analysis of Human Foot Complex in the Sagittal Plane during Level WalkingPLOS ONE, 2013
- A preliminary biomechanical study of a novel carbon–fibre hip implant versus standard metallic hip implantsMedical Engineering & Physics, 2010
- Optimization of Artificial Hip Joint ParametersMaterialwissenschaft und Werkstofftechnik, 2009
- Static and dynamic mechanics analysis on artificial hip joints with different interface designs by the finite element methodJournal of Bionic Engineering, 2007
- Problem of Stress Shielding and Improvement to the Hip Implant Designs: A ReviewJournal of Medical Sciences, 2007
- Static, dynamic and fatigue behavior of newly designed stem shapes for hip prosthesis using finite element analysisMaterials & Design (1980-2015), 2007
- Total hip arthroplasty wear simulation using the boundary element methodJournal of Biomechanics, 2007
- Micromotion, Fit, and Fill of Custom Made Femoral Stems Designed With an Automated ProcessClinical Orthopaedics and Related Research, 1996
- Friction-Induced Vibration, Chatter, Squeal, and Chaos—Part I: Mechanics of Contact and FrictionApplied Mechanics Reviews, 1994
- Stability of press-fit acetabular cupsThe Journal of Arthroplasty, 1992