X-ray Radiation Causes Electromagnetic Interference in Implantable Cardiac Pacemakers
Authors:
- Minoru HiroseMHMinoru Hirose
- Keiichi TachikawaKTKeiichi Tachikawa
- Masanori OzakiMOMasanori Ozaki
- Naoki UmezawaNUNaoki Umezawa
- Toshihiro ShinboTSToshihiro Shinbo
- Hirosuke KobayashiHKHirosuke Kobayashi
- 21 July 2010
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 33 (10) , 1174-1181
Abstract
Background: X-rays are not thought to cause electromagnetic interference (EMI) in implantable cardiac pacemakers. However, x-ray radiation during computed tomography (CT) scanning has been reported to cause EMI in some implantable cardiac pacemakers. The objectives of this study were to identify the location within the pacemakers where x-ray radiation causes EMI and to investigate the association of EMI with the x-ray radiation conditions. Methods: We verified the location where x-ray radiation caused EMI using a CT scanner and conventional radiographic x-ray equipment. An inhibition test and an asynchronous test were performed using five types of implantable cardiac pacemakers. Results: X-ray radiation inhibited the pacing pulses of four types of implantable cardiac pacemakers when the body of each implantable cardiac pacemaker, containing a complementary metal-oxide semiconductor (CMOS), was scanned using a CT scanner. We confirmed that x-ray-induced EMI depends on the x-ray radiation conditions, that is, the tube voltage, tube current, x-ray dose, and direction of x-ray radiation, as well as the sensing thresholds of the implantable cardiac pacemakers. Conclusions: X-ray radiation caused EMI in some implantable cardiac pacemakers, probably because the CMOS component was irradiated. The occurrence of EMI depended on the pacemaker model, sensing threshold of the pacemaker, and x-ray radiation conditions. (PACE 2010; 33:1174–1181)Keywords
This publication has 10 references indexed in Scilit:
- Effects of CT Irradiation on Implantable Cardiac Rhythm Management Devices1Radiology, 2007
- Does High-Power Computed Tomography Scanning Equipment Affect the Operation of Pacemakers?Circulation Journal, 2006
- Radiotherapy to patients with artificial cardiac pacemakersCancer Treatment Reviews, 2005
- Electromagnetic Interference of Implantable Unipolar Cardiac Pacemakers by an Induction OvenPacing and Clinical Electrophysiology, 2005
- The malfunction of permanent pacemaker induced during CT scanningJapanese Journal of Electrocardiology, 2005
- Interactions between pacemakers and security systems.Pacing and Clinical Electrophysiology, 1998
- Interference with Cardiac Pacemakers by Cellular TelephonesNew England Journal of Medicine, 1997
- Pacemaker Failure Induced by RadiotherapyPacing and Clinical Electrophysiology, 1994
- Interference in PacemakersPacing and Clinical Electrophysiology, 1984
- Electromagnetic Interference in Implantable Pacemakers*Pacing and Clinical Electrophysiology, 1978