Comparison of AliveCor KardiaMobile Six-Lead ECG with Standard ECG in Pediatric Patients
- 3 September 2022
- journal article
- research article
- Published by Springer Nature in Pediatric Cardiology
- Vol. 44 (3) , 689-694
Abstract
The AliveCor KardiaMobile (ACKM) is a remote electrocardiogram (ECG) monitoring device. Little research has been conducted on its accuracy with pediatric patients. This prospective study aims to compare the ACKM six-lead device with a standard fifteen-lead ECG in measuring the QTc, QRS, and axis in pediatric patients. Pediatric patients ages 5 to 21 years were enrolled prospectively to have their ECG recorded using an ACKM six-lead device following a recording with the standard 15-lead ECG. A pediatric electrophysiologist measured the QTc, QRS interval, and QRS axis for both ECGs. Bland–Altman analysis was performed to assess agreement among measurements. The study included 141 patients. The mean age was 12.3 ± 4.4 years. Average heart rate was 79 ± 16 bpm. The mean difference in the QTc measurements for a paired standard ECG and ACKM was − 0.6 ms [95% confidence interval − 48 to 47 ms]. Of the ACKM QTc measurements, 117 (83%) were within 30 ms of the standard ECG. The mean difference in paired QRS measurements was − 1.3 ms [95% confidence interval − 23 to 21 ms]. Of the ACKM QRS measurements, 134 (95%) were within 20 ms of the standard ECG. The measured axis was the same for 84% of ACKM and standard ECGs. Over 80% of the ACKM six-lead ECGs produced QTc, QRS, and axis deviation measurements within a clinically useful range of the standard ECG. However, it is not accurate enough to be used consistently in place of a standard ECG for QTc and QRS measurement for pediatric patients.Keywords
All Related Versions
Funding Information
- Scholarly Projects Program, Columbia University Vagelos College of Physicians and Surgeons
This publication has 11 references indexed in Scilit:
- Comparison between a 6‑lead smartphone ECG and 12‑lead ECG in athletesJournal of Electrocardiology, 2021
- Prospective blinded Evaluation of the smartphone-based AliveCor Kardia ECG monitor for Atrial Fibrillation detection: The PEAK-AF studyEuropean Journal of Internal Medicine, 2020
- First real-world experience with mobile health telemonitoring in adult patients with congenital heart diseaseNetherlands Heart Journal, 2018
- Can smartphone wireless ECGs be used to accurately assess ECG intervals in pediatrics? A comparison of mobile health monitoring to standard 12-lead ECGPLOS ONE, 2018
- Assessment of Remote Heart Rhythm Sampling Using the AliveCor Heart Monitor to Screen for Atrial FibrillationCirculation, 2017
- Comparison of QT Interval Readings in Normal Sinus Rhythm Between a Smartphone Heart Monitor and a 12‐Lead ECG for Healthy Volunteers and Inpatients Receiving Sotalol or DofetilideJournal of Cardiovascular Electrophysiology, 2016
- SPEAR Trial: Smartphone Pediatric ElectrocARdiogram TrialPLOS ONE, 2015
- A Smartphone Application to Diagnose the Mechanism of Pediatric Supraventricular TachycardiaPediatric Cardiology, 2015
- QTC intervals can be assessed with the AliveCor heart monitor in patients on dofetilide for atrial fibrillationJournal of Electrocardiology, 2015