Efficacy and safety of flash glucose monitoring in patients with type 1 and type 2 diabetes: a systematic review and meta-analysis
Open Access
- 1 June 2020
- journal article
- review article
- Published by BMJ in BMJ Open Diabetes Research & Care
- Vol. 8 (1) , e001092
Abstract
Introduction Flash glucose monitoring (FGM) is a factory-calibrated sensor-based technology for the measurement of interstitial glucose. We performed a systematic review and meta-analysis to assess its efficacy and safety in patients with type 1 and type 2 diabetes. Research design and methods PubMed, CENTRAL, Scopus and Web of Science were searched in July 2019. Twelve studies with a follow-up longer than 8 weeks, evaluating 2173 patients on prandial insulin, multiple daily insulin injections or continuous subcutaneous insulin infusion were included. The following data were extracted: HbA1c, time in range, time above 180 mg/dL, time below 70 mg/dL, frequency of hypoglycemic events, number of self-monitoring of blood glucose (SMBG) measurements, total daily insulin dose, patient-reported outcomes, adverse events, and discontinuation rate. A comparison with SMBG was conducted. Results FGM use was associated with a reduction in HbA1c (−0.26% (−3 mmol/mol); p=0.002) from baseline to the last available follow-up, which correlated with HbA1c levels at baseline (−0.4% (−4 mmol/mol) for each 1.0% (11 mmol/mol) of HbA1c above 7.2% (55 mmol/mol)). Also, a decrease in time below 70 mg/dL was found (−0.60 hours/day; p=0.04). Favorable findings in patient-reported outcomes and no device-related serious adverse events were reported. When compared with SMBG, FGM was characterized by no statistically different change in HbA1c (p=0.09), with lower number of SMBG measurements per day (−3.76 n/day; pConclusions The present review provides evidence for the use of FGM as an effective strategy for the management of diabetes.Keywords
This publication has 47 references indexed in Scilit:
- Use of Flash Glucose-Sensing Technology for 12 months as a Replacement for Blood Glucose Monitoring in Insulin-treated Type 2 DiabetesDiabetes Therapy, 2017
- A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 DiabetesDiabetes Technology & Therapeutics, 2017
- Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitusBMJ Open Diabetes Research & Care, 2017
- Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled TrialDiabetes Therapy, 2016
- Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trialThe Lancet, 2016
- Real-Time Continuous Glucose Monitoring in Type 1 Diabetes: A Qualitative Framework Analysis of Patient NarrativesDiabetes Care, 2014
- Comparative Effectiveness and Safety of Methods of Insulin Delivery and Glucose Monitoring for Diabetes MellitusAnnals of Internal Medicine, 2012
- The Cochrane Collaboration's tool for assessing risk of bias in randomised trialsBMJ, 2011
- Glycaemic control in type 1 diabetes during real time continuous glucose monitoring compared with self monitoring of blood glucose: meta-analysis of randomised controlled trials using individual patient dataBMJ, 2011
- Relationship of baseline HbA1c and efficacy of current glucose‐lowering therapies: a meta‐analysis of randomized clinical trialsDiabetic Medicine, 2010