An in vitro trial on the effect of arch form on connector size requirements in long span anterior zirconia fixed dental prostheses
Authors:
- 30 September 2022
- journal article
- research article
- Published by Wiley in Journal of Prosthodontics
- Vol. 32 (6) , 534-539
Abstract
Purpose: To test the fracture resistance of maxillary canine to canine fixed partial denture with four missing incisors, with increasing anterior‐cantilevers of the pontics and varying connector sizes.Materials and methods: Two 3D‐printed titanium alloy (Ti6Al4V) models mimicking a maxillary canine to canine fixed partial denture (FPD) with four pontics replacing the incisors were used as master models. Zirconia FPDs were digitally designed and milled with two different connector sizes (9 and 12 mm2) each with three different anterior cantilevers (7, 10, and 13 mm) accounting for 6 test groups. Seven samples were milled for each group generating a total of 42 samples. The zirconia FPDs were cemented on the titanium model using resin modified glass ionomer cement and the model fixated to a variable angle vice. A sinusoidal cyclic waveform load from 50 to 280N was applied using a universal testing machine at a frequency of 30 cycles per second and a total of 5 million cycles.Results: The results of Fisher's exact tests showed that the difference in the proportion of fractured versus nonfractured fixed partial dentures was not statistically significant when comparing the 9 with the 12 mm2 connector size (p = 1.00), as well as when comparing the six test groups (p = 0.2338); on the other hand, it proved to be statistically significant when comparing the 7 mm cantilever with the 10 and 13 mm cantilevers combined (p = 0.0407) indicating that a 7 mm anterior spread of the pontics showed a significantly greater proportion of fixed partial dentures that fractured.Conclusions: Fracture susceptibility was not a function of cantilever length in this testing configuration for anterior FPDs. Retainer crown thickness seems to be a more important parameter than connector size thickness. Based on the results, a smaller connector size (9 mm2) can be used to improve the esthetics of pontics in long span anterior FPDs.Keywords
This publication has 25 references indexed in Scilit:
- Clinical prospective evaluation of zirconia-based three-unit posterior fixed dental prostheses: Up-to ten-year results.Journal of Dentistry, 2016
- Effect of Connector Design on Fracture Resistance in Zirconia-based Fixed Partial Dentures for Upper Anterior RegionThe Bulletin of Tokyo Dental College, 2016
- All-ceramic or metal-ceramic tooth-supported fixed dental prostheses (FDPs)? A systematic review of the survival and complication rates. Part II: Multiple-unit FDPsDental Materials, 2015
- A retrospective evaluation of zirconia-fixed partial dentures in general practices: An up to 13-year studyDental Materials, 2015
- Evaluation of arch width among Class I normal occlusion, Class II Division 1, Class II Division 2, and Class III malocclusion in Indian populationContemporary Clinical Dentistry, 2015
- Load-bearing capacity and the recommended thickness of dental monolithic zirconia single crownsJournal of the Mechanical Behavior of Biomedical Materials, 2014
- State of the art of zirconia for dental applicationsDental Materials, 2008
- Stabilized zirconia as a structural ceramic: An overviewDental Materials, 2008
- In vitro lifetime of dental ceramics under cyclic loading in waterBiomaterials, 2007
- Influence of surface finish and residual stresses on the ageing sensitivity of biomedical grade zirconiaBiomaterials, 2006