We prospectively evaluated the results of our custom cementless femoral stems to ascertain whether this technology produced reasonable clinical function, complication rates, and loosening rates at midterm. Fifty-seven consecutive patients had surgery in 62 hips for primary osteoarthritis at a mean age of 57 years using a three-dimensional computed custom cementless stem. Patients were reviewed at a mean followup of 94.9 months. At review, the mean Harris hip score was 98.8 points (range, 84-100) compared with 61.1 (range, 28-78) points preoperatively. No patient complained of thigh pain. No migration or subsidence was observed. All stems were considered stable according to the radiographic criteria defined by Engh et al. There were no dislocations, no infections, and no reoperations. Our results are comparable with published results from clinical and radiologic points of view. Two problems remain unsolved: the price of a custom stem is twice as expensive as a standard stem; and we need longer term results before definitely recommending this technology as a reasonable alternative to current arthroplasties in younger patients. The data support the continued exploration of this technology with controlled clinical followup. Therapeutic study, Level II-1 (prospective cohort study). See the Guidelines to Authors for a complete description of levels of evidence.Keywords
This publication has 45 references indexed in Scilit:
- Uncemented total hip arthroplasty in patients 50 years of age or youngerThe Journal of Arthroplasty, 1997
- Long-Term Clinical Consequences of Stress-Shielding after Total Hip Arthroplasty without Cement*Journal of Bone and Joint Surgery, 1997
- Hydroxyapatite-Coated Total Hip Femoral Components in Patients Less Than Fifty Years Old. Clinical and Radiographic Results after Five to Eight Years of Follow-up*Journal of Bone and Joint Surgery, 1997
- RADIOLOGICAL AND SURVIVAL COMPARISON OF FOUR METHODS OF FIXATION OF A PROXIMAL FEMORAL STEMThe Journal of Bone and Joint Surgery. British volume, 1997
- Custom total hip arthroplastyThe Journal of Arthroplasty, 1996
- A large collar increases neck resorption in total hip replacement:204 hips evaluated during 5 yearsActa Orthopaedica, 1995
- Pain in the thigh following total hip replacement with a porous-coated anatomic prosthesis for osteoarthrosis. A five-year follow-up study.Journal of Bone and Joint Surgery, 1994
- Thigh pain despite bone ingrowth into uncemented femoral stemsThe Journal of Bone and Joint Surgery. British volume, 1992
- Thigh pain after cementless hip arthroplasty. Annoyance or ill omenThe Journal of Bone and Joint Surgery. British volume, 1992
- Ectopic Ossification Following Total Hip ReplacementJournal of Bone and Joint Surgery, 1973