Profilometer for measuring superfine surfaces
Authors:
- 1 August 2001
- journal article
- Published by SPIE-Intl Soc Optical Eng in Optical Engineering
- Vol. 40 (8) , 1646-1652
Abstract
An interference profilometer for measuring the microprofile of superfine surfaces is described. It has a stabilized He-Ne laser light source with two longitudinal modes and two coaxial interference arms. The measuring arm is focused on the sample surface to form a small spot while the reference arm forms a large spot. As a result, the reference arm provides a constant signal and only the measuring arm will be sensitive to the irregularity of the microprofile. Mechanical vibration has a very small effect on the instrument performance. Even if vibration amplitudes reach 300 nm, the measurement noise is less than 1 nm. The instrument has a vertical resolution better than 0.2 Å root mean square (rms) and smallest form error among existing coaxial profilometers. It is well suited for measuring the microstructure on photoetched surfaces.Keywords
This publication has 6 references indexed in Scilit:
- Investigation of common-path interference profilometryOptical Engineering, 1997
- Differential interference contrast microscope with differential detection for optimizing image contrastApplied Optics, 1996
- Stylus profiling at high resolution and low forceApplied Optics, 1991
- Novel optical approach to atomic force microscopyApplied Physics Letters, 1988
- Optical system for measuring the profiles of super-smooth surfacesPrecision Engineering, 1985
- Stylus profiling instrument for measuring statistical properties of smooth optical surfacesApplied Optics, 1981