MR gradient coil heat dissipation
Authors:
- 1 July 1995
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 34 (1) , 125-132
Abstract
The temperature responses of five different gradient coil designs were modeled using simplified engineering equations and measured. The model predicts that the coil temperature approaches a maximum as an inverse exponential, where the maximum temperature is governed by two parameters: a local power density and a cooling term. The power density term is a function of position and is highest where the current paths have minimum widths and are closely packed. The cooling parameter consists of convective, conductive, and radiative components which can be controlled by (1) providing forced cooling, (2) having a coil former with high thermal conductivity and thin walls, and (3) varying the emissivity of the coil surfaces. For a given gradient strength, the average temperature rise is minimized by designing a coil with a small radius and thick copper. The model predicted the local temperature rise, which is also dependent on the current density, to within 5°C of measured values.Keywords
This publication has 7 references indexed in Scilit:
- A modified quadrupole gradient set for use in high resolution MRI taggingMagnetic Resonance in Medicine, 1994
- Quadrupole gradient coil design and optimization: A printed circuit board approachMagnetic Resonance in Medicine, 1994
- Gradient coil design: A review of methodsMagnetic Resonance Imaging, 1993
- Shielded gradient coils and radio frequency probes for high-resolution imaging of rat brainsMagnetic Resonance in Medicine, 1992
- Coil optimization for MRI by conjugate gradient descentMagnetic Resonance in Medicine, 1991
- Heat Conduction1984
- Improved apparatus for the NMR measurement of self-diffusion coefficients using pulsed field gradientsReview of Scientific Instruments, 1974