Background Saphenous vein graft (SVG) angioplasty using 15-mm articulated, tubular slotted stents results in low (0% to 20%) residual diameter stenoses and infrequent (Methods and Results During a 3-year period, 231 patients with 305 SVG lesions were treated using Palmaz-Schatz coronary (n=108) or biliary (n=123) stents. Cineangiograms were reviewed using qualitative morphological and quantitative angiographic methods. Time-dependent clinical outcome (freedom from death, Q-wave myocardial infarction, or the need for repeat coronary bypass surgery or SVG angioplasty) was assessed using Kaplan-Meier life-table methods. Unstable angina ( P <.001) and recent myocardial infarction ( P =.001) were present more often in patients undergoing biliary stent versus coronary stent placement. Biliary stent–treated SVG lesions were more frequently de novo ( P =.001), ostial in location ( P =.002), ≥10 mm in length ( P =.009), thrombus containing ( P =.001), and ulcerated ( P <.001) than coronary stent–treated SVG lesions. Angiographically, biliary stent–treated lesions had larger reference vessel diameter (3.43±0.59 mm versus 3.10±0.64 mm, P <.001), higher balloon-to-artery ratio (1.15±0.16 mm versus 1.07±0.19, P =.0001), and lower residual diameter stenosis (6±17% versus 14±11% in coronary stent–treated patients; P <.001). Procedural success rates were high (95%), in-hospital major complications were uncommon (Conclusions Despite frequent short-term ischemic syndromes and unfavorable lesion characteristics, both biliary and coronary cohorts have similarly favorable short-term procedural results and long-term clinical outcomes. The increased strut thickness of the biliary stent confers greater fluoroscopic visibility and radial compressive strength in exchange for decreased stent flexibility and added technical demand in stent deployment. Extreme caution is recommended with biliary stent placement in the treatment of SVG lesions as clinical results are highly operator dependent.

This publication has 31 references indexed in Scilit: