Numerical Study of a Customized Transtibial Prosthesis Based on an Analytical Design under a Flex-Foot® Variflex® Architecture
Authors:
- Marco Antonio Hernández-AcostaMHMarco Antonio Hernández-Acosta
- Christopher René Torres-San MiguelCMChristopher René Torres-San Miguel
- Armando Josue Piña-DíazAPArmando Josue Piña-Díaz
- Juan Carlos Paredes-RojasJuan Carlos Paredes-Rojas
- Luis Antonio Aguilar-PerézLuis Antonio Aguilar-Peréz
- Guillermo Urriolagoitia-SosaGUGuillermo Urriolagoitia-Sosa
Open Access
- 22 June 2020
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 10 (12) , 4275
Abstract
This work addresses the design, analysis, and validation of a transtibial custom prosthesis. The methodology consists of the usage of videometry to analyze angular relationships between joints, moments, and reaction forces in the human gait cycle. The customized geometric model of the proposed prosthesis was defined by considering healthy feet for the initial design. The prosthesis model was developed by considering the Flex-Foot® Variflex® architecture on a design basis. By means of the analytical method, the size and material of the final model were calculated. The behavior of the prosthesis was evaluated analytically by a curved elements analysis and the Castigliano theorem, and numerically by the Finite Element Method (FEM). The outcome shows the differences between the analytical and numerical methods for the final prosthesis design, with an error rate no greater than 6.5%.Keywords
This publication has 18 references indexed in Scilit:
- Mechanical and biomechanical analysis of a linear piston design for angular-velocity-based orthotic controlJournal of Rehabilitation Research and Development, 2013
- Muscular strategy shift in human running: dependence of running speed on hip and ankle muscle performanceJournal of Experimental Biology, 2012
- A biomechanical model for encoding joint dynamics: applications to transfemoral prosthesis controlJournal of Applied Physiology, 2012
- Mechanical power and efficiency of level walking with different stride ratesJournal of Experimental Biology, 2007
- Análise biomecânica das articulações do quadril e joelho durante a marcha em participantes idososActa Ortopedica Brasileira, 2007
- Virtual modelling of a prosthetic foot to improve footwear testingProceedings of the Institution of Mechanical Engineers, Part L: Journal of Materials: Design and Applications, 2006
- Load transfer mechanics between trans-tibial prosthetic socket and residual limb—dynamic effectsJournal of Biomechanics, 2004
- Finite element modeling of the contact interface between trans-tibial residual limb and prosthetic socketMedical Engineering & Physics, 2004
- Finite Element Analysis as a Tool for Parametric Prosthetic Foot Design and Evaluation. Technique Development in the Solid Ankle Cushioned Heel (SACH) FootComputer Methods in Biomechanics and Biomedical Engineering, 2003
- Development of a non-linear finite element modelling of the below-knee prosthetic socket interfaceMedical Engineering & Physics, 1995