External compression due to seroma of ventricular assist device outflow graft
- 10 August 2020
- journal article
- case report
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 24 (2) , 258-260
Abstract
A 45-year-old woman with repaired complex congenital heart disease, who underwent placement of Jarvik 2000, a ventricular assist device (VAD) for 4 years, experienced abdominal pain due to outflow graft compression caused by seroma formation between the outflow graft and ringed Gore-Tex graft. We exchanged the pump of Jarvik 2000 and punched several small holes in the new ringed Gore-Tex graft. Seroma formation between the two grafts should be considered as a cause of outflow graft obstruction in patients with the long-term support of VAD, and additional surgical interventions to the ringed Gore-Tex graft may prevent this complication.Keywords
This publication has 7 references indexed in Scilit:
- Successful off-pump surgical release for extrinsic compression of outflow graft obstruction of a left ventricular assist deviceEuropean Journal of Cardio-Thoracic Surgery, 2019
- Left Ventricular Assist Device Outflow Graft Obstruction: A Complication Specific to Polytetrafluoroethylene Covering. A Word of Caution!ASAIO Journal, 2019
- Left ventricular assist device failure due to outflow graft compression by thrombofibrotic exudateThe Journal of Thoracic and Cardiovascular Surgery, 2019
- Late-stage obstruction due to preventative wrapping of left ventricular assist device outflow graftInteractive CardioVascular and Thoracic Surgery, 2019
- A novel external post-pump left ventricular assist device obstructionEuropean Heart Journal - Cardiovascular Imaging, 2019
- Outflow Graft Obstruction Treated With Transcatheter Management: A Novel Therapy for a New DiagnosisThe Annals of Thoracic Surgery, 2017
- Diagnosis and Treatment Algorithm for Blood Flow Obstructions in Patients With Left Ventricular Assist DeviceJACC, 2016