Abstract
Background Clinical translation of the extracorporeal artificial placenta (AP) is impeded by the high risk for intracranial hemorrhage in extremely premature newborns. The Nitric Oxide Surface Anticoagulation (NOSA) system is a novel non-thrombogenic extracorporeal circuit. This study aims to test the NOSA system in the AP without systemic anticoagulation. Methods Ten extremely premature lambs were delivered and connected to the AP. For the NOSA group, the circuit was coated with DBHD-N2O2/argatroban, 100 ppm nitric oxide was blended into the sweep gas, and no systemic anticoagulation was given. For the Heparin control group, a non-coated circuit was used and systemic anticoagulation was administered. Results Animals survived 6.8 ± 0.6 days with normal hemodynamics and gas exchange. Neither group had any hemorrhagic or thrombotic complications. ACT (194 ± 53 vs. 261 ± 86 s; p < 0.001) and aPTT (39 ± 7 vs. 69 ± 23 s; p < 0.001) were significantly lower in the NOSA group than the Heparin group. Platelet and leukocyte activation did not differ significantly from baseline in the NOSA group. Methemoglobin was 3.2 ± 1.1% in the NOSA group compared to 1.6 ± 0.6% in the Heparin group (p < 0.001). Conclusions The AP with the NOSA system successfully supported extremely premature lambs for 7 days without significant bleeding or thrombosis. Impact The Nitric Oxide Surface Anticoagulation (NOSA) system provides effective circuit-based anticoagulation in a fetal sheep model of the extracorporeal artificial placenta (AP) for 7 days. The NOSA system is the first non-thrombogenic circuit to consistently obviate the need for systemic anticoagulation in an extracorporeal circuit for up to 7 days. The NOSA system may allow the AP to be implemented clinically without systemic anticoagulation, thus greatly reducing the intracranial hemorrhage risk for extremely low gestational age newborns. The NOSA system could potentially be applied to any form of extracorporeal life support to reduce or avoid systemic anticoagulation.Keywords
This publication has 37 references indexed in Scilit:
- Effect of Air Exposure and Suction on Blood Cell Activation and Hemolysis in an In Vitro Cardiotomy Suction ModelASAIO Journal, 2013
- A pilot study comparing two polymethylpentene extracorporeal membrane oxygenatorsPerfusion, 2012
- The hemocompatibility of a nitric oxide generating polymer that catalyzes S-nitrosothiol decomposition in an extracorporeal circulation modelBiomaterials, 2011
- Neonatal Outcomes of Extremely Preterm Infants From the NICHD Neonatal Research NetworkPediatrics, 2010
- The attenuation of platelet and monocyte activation in a rabbit model of extracorporeal circulation by a nitric oxide releasing polymerBiomaterials, 2010
- Intracranial hemorrhage in premature neonates treated with extracorporeal membrane oxygenation correlates with conceptional ageThe Journal of Pediatrics, 2004
- More Lipophilic Dialkyldiamine-Based Diazeniumdiolates: Synthesis, Characterization, and Application in Preparing Thromboresistant Nitric Oxide Release Polymeric CoatingsJournal of Medicinal Chemistry, 2003
- The effects of carmeda bioactive surface on human blood components during simulated extracorporeal circulationThe Journal of Thoracic and Cardiovascular Surgery, 1996
- Effect of extracorporeal membrane oxygenation on platelets in newbornsCritical Care Medicine, 1993
- Failure to produce germinal matrix or intraventricular hemorrhage by hypoxia, hypo-, or hypervolemiaExperimental Neurology, 1984