Image quality improvement in optical coherence tomography using Lucy–Richardson deconvolution algorithm
- 6 August 2013
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 52 (23) , 5663-5670
Abstract
Optical coherence tomography (OCT) has the potential for skin tissue characterization due to its high axial and transverse resolution and its acceptable depth penetration. In practice, OCT cannot reach the theoretical resolutions due to imperfections of some of the components used. One way to improve the quality of the images is to estimate the point spread function (PSF) of the OCT system and deconvolve it from the output images. In this paper, we investigate the use of solid phantoms to estimate the PSF of the imaging system. We then utilize iterative Lucy–Richardson deconvolution algorithm to improve the quality of the images. The performance of the proposed algorithm is demonstrated on OCT images acquired from a variety of samples, such as epoxy-resin phantoms, fingertip skin and basaloid larynx and eyelid tissues.Keywords
Funding Information
- NIHR i4i (249889)
- ERC COGATIMABIO (249889)
This publication has 24 references indexed in Scilit:
- Application of regularized Richardson-Lucy algorithm for deconvolution of confocal microscopy imagesJournal of Microscopy, 2011
- Deconvolution methods for image deblurring in optical coherence tomographyJournal of the Optical Society of America A, 2008
- Optical coherence tomographyBritish Journal of Radiology, 2005
- Deconvolution methods for mitigation of transverse blurring in optical coherence tomographyIEEE Transactions on Image Processing, 2005
- Deconvolution in Astronomy: A ReviewPublications of the Astronomical Society of the Pacific, 2002
- Optical coherence tomography (OCT): a reviewIEEE Journal of Selected Topics in Quantum Electronics, 1999
- Image enhancement in optical coherence tomographyusing deconvolutionElectronics Letters, 1997
- Deconvolution and enhancement of optical coherence tomograms1997
- Blind deconvolution by means of the Richardson–Lucy algorithmJournal of the Optical Society of America A, 1995
- Image restoration by Wiener deconvolution in limited-view computed tomographyApplied Optics, 1985