Theoretical and Finite Element Modeling of Fine Kirschner Wires in Ilizarov External Fixator
Authors:
Open Access
- 31 August 2010
- journal article
- Published by ASME International in Journal of Medical Devices
- Vol. 4 (3) , 031001
Abstract
The mechanical behavior of the transosseous elements is a defining factor in the overall stiffness, stability, and reliability of an external fixation system. Mechanics involving the application of thin Kirschner wires in Ilizarov apparatus is yet to be fully explained. To address this problem, load-deflection behavior of the pretensioned thin wires laterally loaded by the bone is necessary to be studied. In this paper, the lateral deflections of thin Kirschner wires are studied both theoretically and computationally. Fully three dimensional finite element (FE) modeling and analyses were performed in which the bone was modeled as a hollow cylinder, and the wire-bone interaction was assumed to be frictionless. The mathematical solution resulted in new exact solutions for the deflection as well as final tension in the wires subjected to the lateral loading under a cylinder representing the bone. Results from the FE analyses turned out to be very close to those from the mathematical solution. The results obtained from theory and FE method are comparable to published experimental findings. Some aspects of the pretensioned thin wire behavior in ring fixation systems, e.g., stiffness-tension proportionality, were revealed in the results. The current study adds to the existing knowledge on the general behavior of tensile elements.Keywords
This publication has 18 references indexed in Scilit:
- Analytical modelling of Kirschner wires in Ilizarov circular external fixator as pretensioned slender beamsJournal of The Royal Society Interface, 2008
- Analytical modelling of Kirschner wires in Ilizarov circular external fixators using a tensile modelProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2008
- Why fine-wire fixators work: An analysis of pressure distribution at the wire–bone interfaceJournal of Biomechanics, 2007
- Avoiding the material nonlinearity in an external fixation deviceClinical Biomechanics, 2004
- Yielding of the clamped-wire system in the Ilizarov external fixatorProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2003
- The effect of clamping a tensioned wire: Implications for the Ilizarov external fixation systemProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2003
- External ring fixators: An overviewProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2000
- Stability of external circular fixation: a multi-variable biomechanical analysisClinical Biomechanics, 1998
- Current Concepts Review - Limb-Lengthening, Skeletal Reconstruction, and Bone Transport with the Ilizarov Method*Journal of Bone and Joint Surgery, 1997
- Comparative study of fracture gap motion in external fixationClinical Biomechanics, 1987