Cylinder mitral and tricuspid valve replacement in neonates and small children
- 26 August 2020
- journal article
- conference paper
- Published by Oxford University Press (OUP) in European Journal of Cardio-Thoracic Surgery
- Vol. 58 (5) , 964-968
Abstract
OBJECTIVES: Atrioventricular valve replacement in small children is associated with high morbidity and mortality. There are no prostheses available with a diameter ˂15 mm. This study reports our initial experience with a cylinder valve for mitral and tricuspid valve replacement in infants and small children.METHODS: Our cylinder valve was hand-made for patients requiring atrioventricuclar valve replacement with an annulus of <15 mm. A 12-mm Contegra valve was prepared and placed inside a 14-mm Gore-Tex tube graft and sutured on both extremities.RESULTS: Eight patients were included, with a median age of 6.9 months (range 1 day to 38 months). Four had mitral and 4 had tricuspid valve replacement. All implants were technically successful, with no significant regurgitation, no stenosis and no left ventricular outflow tract obstruction. There were 3 early deaths from low cardiac output, in patients with significant associated lesions (severe neonatal Ebstein’s, pulmonary artery-intact ventricular septum, biventricular conversion from Norwood stage 1). Two patients required early reintervention: 1 for balloon dilatation for stenosis and 1 for reoperation for paravalvular leak. During follow-up, 2 patients had mitral valve replacement with a 16-mm mechanical valve at 9 and 20 months from the cylinder valve implantation. The remaining 2 patients are alive and well 2 years and 2 months after the procedure.CONCLUSIONS: Cylinder valve replacement of atrioventricular valves was feasible without any technical issues. It was successful in getting out of a difficult situation and allows for somatic growth and implantation of a reasonably-sized mechanical prosthesis on the annulus.Keywords
This publication has 22 references indexed in Scilit:
- Stented bovine jugular vein graft (Melody valve) for surgical mitral valve replacement in infants and childrenThe Journal of Thoracic and Cardiovascular Surgery, 2014
- Annulus upsizing for mitral valve re-replacement in childrenThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Modification of a Stented Bovine Jugular Vein Conduit (Melody Valve) for Surgical Mitral Valve ReplacementThe Annals of Thoracic Surgery, 2012
- Congenital Mitral Valve Stenosis: Anatomic Variants and Surgical ReconstructionSeminars in Thoracic and Cardiovascular Surgery: Pediatric Cardiac Surgery Annual, 2012
- Mitral valve operations at a high-volume pediatric heart center: Evolving techniques and improved survival with mitral valve repair versus replacementAnnals of Pediatric Cardiology, 2012
- Short-Term Performance of the Transcatheter Melody Valve in High-Pressure Hemodynamic Environments in the Pulmonary and Systemic CirculationsCirculation: Cardiovascular Interventions, 2011
- Mitral valve replacement in infants and children 5 years of age or younger: Evolution in practice and outcome over three decades with a focus on supra-annular prosthesis implantationThe Journal of Thoracic and Cardiovascular Surgery, 2008
- Long-term follow-up after mitral valve replacement in childhood: poor event-free survival in the young child☆European Journal of Cardio-Thoracic Surgery, 2007
- When Reconstruction Fails or is Not Feasible: Valve Replacement Options in the Pediatric PopulationSeminars in Thoracic and Cardiovascular Surgery: Pediatric Cardiac Surgery Annual, 2007
- Mitral and tricuspid valve anular diameter in normal children determined by two-dimensional echocardiographyThe American Journal of Cardiology, 1985