A new approach to electron‐beam reference dosimetry
- 1 March 1998
- journal article
- Published by Wiley in Medical Physics
- Vol. 25 (3) , 310-320
Abstract
A new approach is proposed for electron‐beam dosimetry under reference conditions and data necessary to use this approach are presented. The approach has the following features: it uses ion chambers and starts from an absorbed‐dose calibration factor for to be consistent with the present proposal for the new AAPM photon‐beam protocol; it uses to specify the beam quality and the reference depth, (all quantities in cm), recommended by Burns et al. [Med. Phys. 23, 383–388 (1996)]; it has a formalism which is parallel to the formalism for photon‐beam dosimetry; it fully accounts for the impact on stopping‐power ratios of realistic electron beams; it allows an easy transition to using primary standards for absorbed dose to water in electron beams when these are available. The equation for dose to water under reference conditions is: The term is not needed with plane‐parallel chambers but corrects for gradient effects with cylindrical chambers and is measured in the user's beam. The parameter is associated with converting the absorbed‐dose calibration factor into one for an electron beam of quality and contains most of the chamber to chamber variation. Calculated values of are presented as well as Monte Carlo calculated values for plane‐parallel chambers in a water phantom irradiated by a beam since these are needed to calculate The factor is a function of and converts the absorbed‐dose calibration factor to that for the electron‐beam quality of interest. Two analytical expressions are presented which are close to universal expressions for all cylindrical Farmer‐like chambers and for well‐guarded plane‐parallel chambers respectively. Calculated values are presented graphically for electron beams with energies between 5 and 50 MeV.Keywords
This publication has 19 references indexed in Scilit:
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- Mean energy, energy‐range relationships and depth‐scaling factors for clinical electron beamsMedical Physics, 1996
- Calculation of stopping‐power ratios using realistic clinical electron beamsMedical Physics, 1995
- The calibration and use of plane‐parallel ionization chambers for dosimetry of electron beams: An extension of the 1983 AAPM protocol report of AAPM Radiation Therapy Committee Task Group No. 39Medical Physics, 1994
- Verification of absorbed doses determined with thimble and parallel‐plate ionization chambers in clinical electron beams using ferrous sulphate dosimetryMedical Physics, 1994
- Effect of size and composition of the central electrode on the response of cylindrical ionization chambers in high-energy photon and electron beamsPhysics in Medicine & Biology, 1993
- Absorbed dose beam quality factors for the dosimetry of high-energy photon beamsPhysics in Medicine & Biology, 1992
- The advantages of absorbed‐dose calibration factorsMedical Physics, 1992
- Clinical electron‐beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25Medical Physics, 1991
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983