A mock circulation model for cardiovascular device evaluation
Authors:
- K A M A PenningsKPK A M A Pennings
- M J G van de MolengraftMMM J G van de Molengraft
- N H J PijlsNPN H J Pijls
- F N van de VosseFVF N van de Vosse
- M C M RuttenMRM C M Rutten
- 12 March 2014
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 35 (4) , 687-702
Abstract
The aim of this study was to develop an integrated mock circulation system that functions in a physiological manner for testing cardiovascular devices under well-controlled circumstances. In contrast to previously reported mock loops, the model includes a systemic, pulmonary, and coronary circulation, an elaborate heart contraction model, and a realistic heart rate control model. The behavior of the presented system was tested in response to changes in left ventricular contractile states, loading conditions, and heart rate. For validation purposes, generated hemodynamic parameters and responses were compared to literature. The model was implemented in a servo-motor driven mock loop, together with a relatively simple lead-lag controller. The pressure and flow signals measured closely mimicked human pressure under both physiological and pathological conditions. In addition, the system's response to changes in preload, afterload, and heart rate indicate a proper implementation of the incorporated feedback mechanisms (frequency and cardiac function control). Therefore, the presented mock circulation allows for generic in vitro testing of cardiovascular devices under well-controlled circumstances.Keywords
This publication has 34 references indexed in Scilit:
- Rapid prototyping compliant arterial phantoms for in-vitro studies and device testingJournal of Cardiovascular Magnetic Resonance, 2013
- Intracardiac Visualization of Transcatheter Aortic Valve and Valve-in-Valve Implantation in an In Vitro Passive Beating HeartJACC: Cardiovascular Interventions, 2013
- Effects of Intra‐Aortic Balloon Pump Timing on Baroreflex Activities in a Closed‐Loop Cardiovascular Hybrid ModelArtificial Organs, 2012
- A Mock Circulatory System With Physiological Distribution of Terminal Resistance and Compliance: Application for Testing the Intra‐Aortic Balloon PumpArtificial Organs, 2010
- A Modified Elastance Model to Control Mock Ventricles in Real-Time: Numerical and Experimental ValidationASAIO Journal, 2008
- Left Ventricular Assist Device and Drug Therapy for the Reversal of Heart FailureNew England Journal of Medicine, 2006
- Dependence of Intramyocardial Pressure and Coronary Flow on Ventricular Loading and Contractility: A Model StudyAnnals of Biomedical Engineering, 2006
- A Complete Mock Circulation Loop for the Evaluation of Left, Right, and Biventricular Assist DevicesArtificial Organs, 2005
- Introducing a baroreflex model for studying cardiovascular effects of mental workloadPsychophysiology, 2004
- An Optimal Estimation Method for Nonlinear Models of Mechanical SystemsJournal of Dynamic Systems, Measurement, and Control, 1994