Graphene Quantum Dots and Enzyme-Coupled Biosensor for Highly Sensitive Determination of Hydrogen Peroxide and Glucose
Open Access
- 7 June 2018
- journal article
- research article
- Published by MDPI AG in International Journal of Molecular Sciences
- Vol. 19 (6) , 1696
Abstract
In this paper, a simple and specific graphene quantum dots (GQDs)-based fluorescent biosensor adopted for the determination of glucose based on the combination of the enzyme-coupled method and fluorescence quenching mechanism is demonstrated. Glucose was oxidized by the enzyme glucose oxidase (GOx), forming hydrogen peroxide (H O ) via the catalysis by horseradish peroxidase (HRP). H O was then employed to oxidize phenol to quinone, which led to effective quenching effect in the GQDs–GOx–HRP–phenol system. By optimizing the reaction conditions of the GQDs-enzyme system, a linear relationship between the concentration of glucose and the fluorescence intensity over a range of 0.2–10 mol/L was obtained. The limit of detection for glucose is 0.08 mol/L. The present biosensor for the determination of glucose showed satisfactory reproducibility and accuracy in human serum samples. Since the enzymes have high specificity and unique affinity to the certain substance, the enzyme-coupled system promises a sensitive way for further detection of those chemicals which could be oxidized by enzymes and generated H O or glucose. GQDs and other fluorescent materials coupled with several enzymes can be applied to extensive sensing field.
Keywords
Funding Information
- National Natural Science Foundation of China (31772058)
- Natural Science Foundation of Jilin Province (20160520127JH)
- China Postdoctoral Science Foundation (2016T90260, 2013M541309)
This publication has 55 references indexed in Scilit:
- Self-Powered Implantable Skin-Like Glucometer for Real-Time Detection of Blood Glucose Level In VivoNano-Micro Letters, 2018
- Biosensing platform for the detection of uric acid based on graphene quantum dots and G-quadruplex/hemin DNAzymeAnalytica Chimica Acta, 2017
- The in vitro and in vivo toxicity of graphene quantum dotsBiomaterials, 2014
- Fluorescent blood glucose monitor by hemin-functionalized graphene quantum dots based sensing systemAnalytica Chimica Acta, 2014
- Graphene Quantum DotsParticle & Particle Systems Characterization, 2013
- Glucose oxidase-functionalized fluorescent gold nanoclusters as probes for glucoseAnalytica Chimica Acta, 2013
- Assembly of Anionic Conjugated Polymer with 6‐O‐Modified PNP‐β‐Galactoside for Fluorescence Logic‐signal‐based Multiplex Detections of EnzymesMacromolecular Rapid Communications, 2010
- Graphene Oxide: Intrinsic Peroxidase Catalytic Activity and Its Application to Glucose DetectionAdvanced Materials, 2010
- Review: Carbon nanotube based electrochemical sensors for biomoleculesAnalytica Chimica Acta, 2010
- Nanomedicine and its potential in diabetes research and practiceDiabetes/Metabolism Research and Reviews, 2008