Performance of the Iris iQ ® 200 Elite analyser in the cell counting of serous effusion fluids and cerebrospinal drainage fluids
- 11 August 2011
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 64 (12) , 1123-1127
Abstract
Aims: Evaluation of the Iris iQ ® 200 Elite analyser, initially designed for urinary cell counting, for the analysis of biological fluids (serous effusion fluids and cerebrospinal drainage fluids) and comparison of its performance with that of the manual microscopic method. Methods: Routine samples (ascite fluids, pleural fluids and cerebrospinal fluids) were evaluated in terms of red blood cells and nucleated elements using the iQ ® 200 analyser and the manual method. The authors compared the reliability, repeatability and speed of the two techniques. In addition, the authors assessed the contribution of two different sample dilution processes to the improvement of iQ ® 200 analyser cytological results. Results: Very good agreements were found between the two methods and between the two sample dilution processes. Regarding the repeatability, the coefficients of variation obtained with the iQ200 were slightly higher than those obtained by the manual method. Besides, the difference in the speed of the two methods was not significantly different for series with ® 200 Elite analyser has allowed us to obtain reliable results, equivalent to that of the manual method, for cell enumeration in biological fluids. Although the speed of this instrument needs to be improved for larger series of samples, it enables standardised and objective cytological results to be obtained and represents an alternative to the usual manual microscopic method. Moreover, automation of such analyses permits saving of technician time.Keywords
This publication has 13 references indexed in Scilit:
- Comparative Evaluation of the Iris iQ200 Body Fluid Module With Manual Hemacytometer CountAmerican Journal of Clinical Pathology, 2009
- The Clinical Reliability of Automated Cerebrospinal Fluid Cell Counts on the Beckman-Coulter LH750 and Iris iQ200American Journal of Clinical Pathology, 2009
- Evaluation of iQ200 Automated Urine Microscopy AnalyzerAnnals of Laboratory Medicine, 2008
- A Multicenter Evaluation of the Iris iQ200 Automated Urine Microscopy Analyzer Body Fluids Module and Comparison With Hemacytometer Cell CountsAmerican Journal of Clinical Pathology, 2008
- Risk factors for infections related to external ventricular drainageActa Neurochirurgica, 2008
- An Evaluation of the Cell-Dyn 3200 for Counting Cells in Cerebrospinal and Other Body FluidsLaboratory Hematology, 2005
- Analytic Performance of the iQ200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell ChambersAmerican Journal of Clinical Pathology, 2005
- An Evaluation of the Utility of Performing Body Fluid Counts on the Coulter LH 750Laboratory Hematology, 2004
- Automated Flow Cytometric Analysis of Blood Cells in Cerebrospinal FluidAmerican Journal of Clinical Pathology, 2004
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986